electrical connections
1. Ensure that the fl ow sensor frame is connected to ground (
⏚)
; it is recommended that the ground
terminal at the outside of the sensor is used for this purpose. This will prevent against undesirable
interference, particularly for detached versions when shielding the link cable between the sensor
and electronic unit while connected to the frame.
Remove the upper cover of the instrument to gain access to terminals for the power cable and
the signal cables.
CAUTION
PLEASE BE AWARE THAN AN INDUCTION FLOW METER IS AN ELECTRONIC DEVICE;
THE FOLLOWING STEPS SHOULD BE PERFORMED BY A QUALIFIED ELECTRICIAN!
TERMINAL BLOCKS
Terminal
Description
FLOW C
tube flooding indication electrode
FLOW A
measuring electrode
FLOW B
measuring electrode
FLOW
⏚
ground terminal of the sensor
PWR D
solenoid
PWR E
solenoid
BIN1
programmable multi-functional output
BIN2
programmable multi-functional output
ANAL
programmable analog output
FREQ
frequency output (for calibration only)
230 V L
phase conductor 230V / 60 Hz
230 V PE
protective conductor
230 V N
neutral conductor 230V / 60 Hz
AC/DC L
supply con/-
AC/DC PE
protective conductor
AC/DC L
supply con/-
PO1
protective fuse 100 mA design with 230 V/50 Hz
supply protective fuse 1A design with AC/DC supply
COMPACT VERSION
REMOTE VERSION
Control module
Sensor