www.h2flow.net
•
Tel: 888-635-0296 (Toll Free)
OR
(+1) 419-841-7774 (International)
1.
description
The induction FM500 is designed for measuring both clean and waste water in water management applications.
The product is characterized by high accuracy and stability of flow rate measurement in whole range of flow rates from 0.33
to 33 ft/s in both directions
for
liquids ≥ 10µs/cm.
Measured values are displayed on a large graphical display, which is a standard component of the instrument as well as a
keyboard and three programmable isolated outputs for remote transmission of measure values. Two of these outputs can be
used as pulse or status outputs and the third one as an active current output. The standard delivery also includes an internal
data logger and RS232 communication port.
A GSM module for remote data reading via SMS text message can be ordered as an option.
Setting of flow meter parameters is performed using the keyboard and intuitive display menu or using service software via the
RS232 communication port.
3
2.
flow ranges
GPM
PIPE DIAMETER
(inches)
1”
2”
3”
4”
6”
8”
10”
12”
Minimum
0.79
3.17
7.93
12.68
28.53
50.72
79.25
111
Maximum
79
317
793
1,268
2,853
5,072
7,925
11,095
14”
16”
20”
24”
28”
32”
36”
40”
Minimum
152
198
317
444
610
793
1,007
1,244
Maximum
15,216
19,813
31,700
44,380
61,023
79,251
100,649
124,425
NOTE:
Please consult the factory for accurate flow range data for models smaller than 1-inch in diameter and larger than