G1.
Datalogger setting without time mask
1. Use the cursor keys to select the option
DATALOGGER
from the main menu and confirm your selection by pressing the
EXE
button. A sub-menu will be displayed.
2. Select the option
MASK
and confirm your selection by pressing the
EXE
button.
3. A time mask is displayed. The time mask is disabled if it is in format xx - xx - xx xx : xx (day - month - year hrs. : mins.).
If any position shows a character other than ‘x’, select the
EDIT
command and use the cursor keys to correct it to x. Use
the command
RETURN
to return to the sub-menu.
4. Select the
VAR
option and confirm your selection by pressing the
EXE
button. In the menu, select one of the optional
values to be recorded (immediate flow rate Q, maximum flow rate during chosen interval MAX, minimum flow rate during
chosen interval MIN, or average flow rate during chosen interval AVERAGE). The selected option will be marked with an
asterisk ( * ). Use the
RETURN
command to return to the sub-menu.
5. Select the option
PER
and confirm your selection by pressing the
EXE
button. Use the
EDIT
command to enable interval
editing in minutes. Enter the interval length and use the
RETURN
command to return to the sub-menu.
6. Select
START
. The
RUN
option should be marked with an asterisk *; this means that the datalogger is operating.
7. Select the
RETURN
option to complete the set up.
The instrument will ask if you want to save the new parameters by displaying the text
SAVE SETUP?
. Select
YES
to confirm
and the flow meter will save the new parameters and switch to normal display mode. If
NO
is selected, parameters will not be
saved.
The instrument will switch to display mode and work with the new parameters until it is switched off. When the instrument is
switched on again, it will use the original/default parameters.
G2.
Datalogger setting with time mask (using real time)
1. Use the cursor keys to select the option
DATALOGGER
from the main menu and confirm your selection by pressing the
EXE
button. A sub-menu will be displayed.
2. Select the option
MASK
and confirm your selection by pressing the
EXE
button.
3. A time mask is displayed. The time mask is disabled if it is in format xx - xx - xx xx : xx (day - month - year hrs. : mins.).
Select the
EDIT
command and use the cursor keys to set the required time mask format. Use the command
RETURN
to
return to the sub-menu.
4. Select the
VAR
option and confirm your selection by pressing the
EXE
button. In the menu, select one of the optional
values to be recorded (immediate flow rate Q, maximum flow rate during chosen interval MAX, minimum flow rate during
chosen interval MIN, or average flow rate during chosen interval AVERAGE). The selected option will be marked with an
asterisk ( * ). Use the
RETURN
command to return to the sub-menu.
5. Select the option
PER
and confirm your selection by pressing the
EXE
button. Use the
EDIT
command to enable interval
editing in minutes. Enter the interval length and use the
RETURN
command to return to the sub-menu.