technical data
file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc
© 2013 h/p/cosmos sports & medical gmbh author: fh [email protected] created 07.02.2013 printed 04.03.2013 page: 148 of 197
11.2
Running machines, 170-190/65 (quasar / pulsar)
stellar
®
, quasar
®
, pulsar
®
locomotion
®
190/65 med
h/
p/
cosmos st
ellar
h/
p/
cosmos st
ellar lt
h/
p/
cosmos quas
ar
h/
p/
cosmos quas
ar lt
h/
p/
cosmos st
ellar m
ed
h/
p/
cosmos st
ellar m
ed lt
h/
p/
cosmos quas
ar med
h/
p/
cosmos quas
ar med lt
h/
p/
cosmos pulsar
h/
p/
cosmos pulsar lt
h/
p/
cosmos locomot
ion 190/
65 E med
h/
p/
cosmos locomot
ion 190/
65 DE med
UserTerminal
(display & keyboard)
6 LCD displays / 6 keys / display of mode and unit with LED.
“lt-models“ have no UserTerminal (no display, no keyboard). Control via interface.
E: no Terminal
DE: TouchScreen
Power supply AC
(standard, read nameplate),
dedicated line required.
U: 100 V ~ / 110 V ~ / 200 V ~ / 208 V ~ / 230 V ~
including N and PE
f: 50/60 Hz
Fuse (standard, read nameplate)
C 16 A at 230 V ~
Power input (long time)
1600 VA
Power input (momentary)
3500 VA
(peak performances for milliseconds; max. 1s/min during heel strike)
Drive motor capacity
3300 VA
Elevation motor capacity
not applicable
500 VA
not applicable
500 VA
Safety standards
c
IEC EN 60335-1(VDE 0700)
IEC EN 60601-1-2,
IEC EN 60601-1-4, VDE 0701,
EN 957-1, EN 957-6, Machinery
Directive 2006/42/EC
n
MDD, directive 93/42/EEC + 2007/47/EC,
IEC EN 60601-1 (VDE 0750), VDE 0751,
IEC EN 60601-1-2, IEC EN 60601-1-4,
EN 957-1, EN 957-6, Machinery Directive 2006/42/EC
Leakage current
0.6 mA
0.2 mA
Isolation transformer
---
2000 VA
Degree of protection
class I
M
/ --- / IP 20
class I
M
/ type B
R
/ IP 20
Overvoltage category
Life-Earth < 300 V
transient over-voltage limit 1500 V
Category III: Life-Earth < 300 V
transient over-voltage limit 4000 V
Mode of operation
continuous operation with intermittent loading (according to IEC 60601-1)
Field of application / accuracy
sports and fitness: S,I,A (EN957)
sports and medical: S,I,A (EN957)
Classification according to MDD
---
active therapeutic device and active diagnostic device, risk class
II b
Speed
Definition
0.0 ... 25.0 km/h
0.1 km/h
0.0 ... 40.0 km/h
0.1 km/h
0.0 … 10.0 km/h
0.1 km/h
Speed accuracy*
+/- 5 %
Option special speed
0.0 ... 10.0 km/h, 0.0 ... 40.0 km/h, 0.0... 45.0 km/h
0.0 ... 10.0 km/h
0.0... 45.0 km/h
0.0 ... 25.0 km/h
Angle of inclination**
Definition
not applicable
0.0 ... 28.0 %
0.1 %
not applicable
0.0 ... 28.0 %
0.1 %
- 25.0 % ... + 25.0 %
0.1 %