page 71 of 99
cos105000_170-190_65, rev. 1.03
© 1988 - 2019 h/p/cosmos sports & medical gmbh
Option Description
Comment / Display
OP28 Maximum acceleration and deceleration level
The selected maximum level is valid for all
acceleration and deceleration processes in all
modes and profiles
flashes, reports the max. acceleration / deceleration for all
modes and profiles (standard: level 4)
The maximum acceleration and deceleration level is NOT valid
for control and operation via V24/RS232 interface. In this case
the acceleration and deceleration level is set in option 29 or in the
corresponding command of the h/p/cosmos coscom protocol.
OP29 Standard acceleration and deceleration level for
RS 232 interface
The selected acceleration and deceleration level is valid for control
and operation via RS232 interface. This option is very helpful if the
peripheral equipment (e.g. ECG, ergospirometry, PC) does not offer a
menu for acceleration and deceleration levels.
flashing, reports: 1 ... 5, (standard: 1) for the acceleration and
deceleration level for all speed commands via RS 232. The maximum
adjustable value depends on the setting of option 28.
Note: If the peripheral equipment sends an acceleration and
deceleration command via the h/p/cosmos coscom protocol, the
selected level in option 29 or in the corresponding command of the
h/p/cosmos coscom protocol.
OP40 Locking and unlocking of the treadmill
OFF = After switching on, the treadmill is completely locked / not
accessible. To unlock the treadmill, press the b, - and START
simultaneously.
While locked the display shows "no ACCESS"
ON = treadmill is unlocked / accessible (standard)
OP41 Locking and unlocking the manual mode
OFF = manual mode is locked / not accessible
ON = manual mode is unlocked / accessible (standard)
OP42 Locking and unlocking the profile mode
OFF = profile mode is locked / not accessible
1 ... 6 = profile mode is unlocked / accessible up to the selected profile
number standard: 6
Example: Selected profile number = 3: The profiles 1-3 can be
selected, the profiles 4 – 6 cannot be selected
OP43 Locking and unlocking the cardio mode
OFF = cardio mode is locked / not accessible
ON = cardio mode is unlocked (standard)
OP44 Locking and unlocking the test mode
OFF = test mode is locked / not accessible
1 ... 94 = test mode is unlocked / accessible up to the selected test
profile number standard: 24
selected test number = 5: The test profiles 1-5 can be selected, the
test profiles 6 – 94 cannot be selected
OP45 Report mode display “Index“
0 = Display alternates (default)
1 = Display permanently in MET
2 = Display permanently in kJ
3 = Display permanently in Watt