SECTION 1 INTRODUCTION
Gyrus ACMI PlasmaKinetic SuperPulse Generator (Endourology)
USER MANUAL
Part Number: 144020-LB
SuperPulse Generator will automatically select a default power and mode setting according to the
instrument type. Power and mode adjustments can be performed by the front panel buttons with the
selected output shown on the user display.
Instruments are supplied in a sterile, single use format and connect to the SuperPulse Generator via a
limited re-use connector cable. Activation of the electrosurgical current is by means of a footswitch,
attached via the back panel. Activation is inhibited by safety circuitry until both instrument and
connector cable have been properly coupled to the SuperPulse Generator. An audible alarm will
sound whenever electrosurgical energy is being output. Diagnostic circuits within the SuperPulse
Generator continuously monitor system performance. Any detected faults are indicated as symbols on
the user display in conjunction with the illumination of the front panel warning symbol.
1 C. Indication for Use
The Gyrus ACMI PK Superpulse System is intended for use for ablation, removal, resection and
coagulation of soft tissue and where associated haemostasis is required in open, endoscopic and
laparoscopic surgical procedures.
The device is intended for use by qualified medical personnel trained in the use of electrosurgical
equipment.
1D. Contraindications
for
Use
The use of this device is contraindicated in patients with the following conditions:
Carcinoma of the bladder or prostate without tissue diagnosis.
Patients with urinary tract infection.
Patients with incipient renal failure.
It is advisable to monitor the input and output volumes of the irrigation fluid in all patients but
especially those with cardiovascular insufficiency or poor renal function.
Patients with Pacemakers
Use with caution in the presence of internal or external pacemakers. Interference from an
electrosurgical current can cause a pacemaker to enter an asynchronous mode or can block the
pacemaker effect entirely. For further information, consult the pacemaker manufacturer or hospital
Cardiology Department.
1E. System
Description
The SuperPulse Endourology System (figure 1.1) is designed for resectoscopic and endoscopic
urological procedures performed on the lower urinary tract. A typical system comprises the following
items:
A SuperPulse Generator (Model 744000)
A Gyrus ACMI Footswitch (Model 744010)
Suitable connector cable to connect to the Gyrus or Olympus Resectoscope.
Use only approved accessories - Contact your sales representative.
When used with a PlasmaKinetic
Resectoscope
A suitable Gyrus ACMI Urology Electrode (e.g. PK Plasma Sect, PK SuperSect, Superloop
and PK Button) (see Fig 1.2)
When used with a suitable urological endoscope (e.g. Cystoscope)
A PK Plasma-Cise
or PK Plasma Cut
instrument (see Fig 1.3)
When used with an approved Resectoscope
Olympus Electrodes – Compatible with the Gyrus ACMI SuperPulse System.
Supplied with connector cable
Page 1-2
SECTION 1
INTRO
D
UCTION
Summary of Contents for PlasmaKinetic
Page 2: ......