SECTION 14 EMC TABLES
Gyrus ACMI PlasmaKinetic SuperPulse Generator (Endourology)
USER MANUAL
Part Number: 144020-LB
Recommended Separation Distances between Portable and Mobile RF Communications Equipment
and the SuperPulse Generator
The SuperPulse Generator is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the SuperPulse Generator can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the SuperPulse Generator as recommended below, according
to the maximum output power of the communications equipment.
Separation Distance According to Frequency of Transmitter
m
Rated Maximum
Output Power of
Transmitter
W
150 kHz to 80 MHz
d
= [3.5/V
1
]
√
P
80 MHz to 800 MHz
d
= [3.5/E
1
]
√
P
800 MHz to 2.5 GHz
d
= [7/E
1
]
√
P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17
1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Page 14-4
SECTION 14
EMC TABLES