PPH-1503 User Manual
118
Status Commands
Command
:STATus:PRESet
Function
Resets the unit to the default settings. Clears the
operation event enable register, measurement
event enable register and questionable event
enable register to return to the default status.
Example
:STATus:PRESet
Command
:STATus:OPERation[:EVENt]?
Function
Read the operation event register.
Example
:STATus:OPERation?
Reads the operation event register.
Command
:STATus:OPERation:CONDition?
Function
Read the operation condition status register.
Example
:STATus:OPERation:CONDition?
Returns the contents of the operation condition
status register.
Command
:STATus:OPERation:ENABle <NRf>
Function
Programs the operation enable status register.
Description
<NRf>
8: CL (Current enable bit).
16: CLT (Current limit tripped enable
bit).
64: PSS (Power supply shutdown
enable bit).
Example
:STATus:OPERation:ENABle 64
Enable the the power supply shutdown bit.