PEL-3000 Series Parallel Assembly Guide
12
The current sink power of PEL-3111H and PEL-3211H is 1KW:2KW;
therefore, the ratio of current sink will always be the exactly 1:2.
E.g., PEL-3322H (3KW); when current sink is set 30A, Master (PEL-3111H)
will sink current by 10A, whilst Booster (PEL-3211H) will sink current by
20A.
E.g., PEL-3955H (9KW); when current sink is set 90A, Master
(PEL-3111H) will sink current by 10A, whilst the rest of the 4 Booster
(PEL-3211H) will sink current by 20A individually.
First dismantle all of the positive parallel copper sticks and parallel cables
in the rear output terminal of unit.
Check the 1st Booster current:
I.
Connect the parallel cable to the J2 connector in the rear of
PEL-3111H in one end and the J1 connector in the rear of
PEL-3211H in the other end.
II.
Power on both PEL-3111H and the 1st PEL-3211H, while keep
the rest units power-off. Set the unit PEL-3111H as the parallel
mode via the following procedure:
III.
Select from the main interface
a.
Configure > Next Menu > Parallel > Operation; select: Master
b.
Configure > Next Menu > Parallel > Parallel; select: OFF
c.
Configure > Next Menu > Parallel > Booster; select: 1