5. APPLICATION
5-1 The Table of the error message
Code
Cause
Resolvent
Err-01
Input voltage over 60V
Lower the input voltage
Err-02
Input power exceeds setting
Set higher power value or lower the voltage and
current
Err-03
Input voltage and power exceed setting
Lower the input voltage and set a higher power
value
Err-04
Input current exceeds setting
Lower the input current
Err-05
Input voltage and current exceed setting
Same as Err-01 and Err-04
Err-06
Input current and power exceed setting
Same as Err-02 and Err-04
Err-07
Input voltage, current and power exceed setting Same as ERR-01, Err-02 and Err-04
Err-08
Error of the adjusting data when performing
CHECKSUM
Send the instrument back to the manufacturer
for adjustment or adjust by yourself.
Err-09
Error inaccurate when performing CHECKSUM
Send the instrument back to the manufacturer
for repair
Err-10
The time of running AUTO has not been set
Re-set the time of execution
Err-11
Key-in value exceeds input range
Key in the value again
Err-12
The setting of STEP exceeds the resolution of
the range between chosen
Set the value of STEP again
Err-13
The setting of STEP exceeds output range
Set the value of V.I.R. again
To clear the Error Message: [SHIFT] + [C.C. MODE]