4-8 Calibrating the D/A, A/D in 0.6A range of C.C. MODE
n
Set voltage of POWER SUPPLY to 60V and current of it to 0.8A.
o
Connect the output terminal of POWER SUPPLY to a DMM in 2A range in series. Hook up the DMM
with the input terminal of the electronic load.
p
Test the input terminal of the load by the 200V range of a DMM. Press SHIFT and C.V. MODE, then key in
3, 8, 0, 7, and
↵
.
q
Adjust the POWER SUPPLY until t he DMM display reads 60.00V.
r
When MEMORY indicates “01” and A indicates “CL07”, key in the current reading (no less than 6 digits
including decimal) of the DMM (in unit of “A”) and press
↵
.
s
When MEMORY indicates “02”, key in the current reading (no less than 7 digits including decimal) of the
DMM (in unit of “A”) and press
↵
.
Note:
Applied instruments -- POWER SUPPLY (60V, 6A); DMM (5 1/2 digits)
Applied situation --The readings of C.C. MODE in 0.6A range are not reliable.
4-9 Calibrating the D/A in 0.6A range of W SET
n
Set voltage of POWER SUPPLY to 60V and current of it to 0.8A.
o
Connect the output terminal of POWER SUPPLY to a DMM in 2A range in series. Hook up the DMM
with the input terminal of the electronic load.
p
Test the input terminal of the load by the 200V range of a DMM. Press SHIFT and C.V. MODE, then key in
3, 8, 0, 8, and
↵
.