OPERATION
81
Defined Settings
The DEF1 and DEF2 settings are user defined
settings. By default they are set to 50.00 Hz and
60.00 Hz, respectively. The MAX and MIN soft-
keys set the frequency to the maximum and
minimum, respectively.
4.
Repeat the previous steps 1 ~ 2 to set frequency
with the scroll wheel.
5.
Press and hold the
DEF1
or
DEF2
soft-key until
“Saved to DEF1/2” is displayed. This will save
the frequency setting to the DEF1 or DEF2 soft-
key individually.
Example of
Frequency Setting
in AC+DC-INT
Frequency setting
Defined setting
F3
F4
F1
F2
Example of Signal
Setting in the
AC+DC-EXT
Signal setting
Note
Trying to set the frequency outside of the frequency
limit will result in a frequency setting error being
displayed on the screen.
FREQ setting under each output mode has its own
DEF1 and DEF2 saved values, respectively.