TEST MODE FUNCTION
159
Secondary
settings
CT (Constant), KP (Keep), SP
(Sweep)
Note: Step 0 can only be set to
either CT or SP.
10.
Go to the
Wave
setting and choose which
waveform to output.
Wave
SIN, SQU, TRI, ARB1 - 16
11.
Go to the
Jump To
setting and choose which
step to jump to, or turn the setting off.
Step
ON, OFF, 0 ~ 999
12.
Go to the
Jump Cnt
setting and set the number
of times the current step will loop.
Jump Cnt
1 ~ 9999, 0
Note: A setting of 0 will set the
number of jump step to be infinite.
13.
Go to the
Branch 1/2
setting and set a step to
branch to.
Branch 1, 2
ON, OFF, 0 ~ 999
14.
Go to the
Term
setting and set the step
termination setting. CONTI will automatically
go to the next step at the end of the step. END
will return to step 0. HOLD will stay at the
current step until you allow the sequence to
continue to the next step.
Term
CONTI, END, HOLD
15.
Go to the
Sync Code
setting and set the
synchronous code when the step has started.