ASR-3000 Series User Manual
152
Step
Assigns the step number.
Time
Sets the step duration time. This step
time is exclusive of any transition
time needed to match start phases
and stop phase. See the diagram on
page 155 for details.
ACV
Sets the AC voltage level. There are 3
secondary voltage settings that
determine how the voltage is output.
CT: Sets the voltage level of the step
immediately to ACV values.
KP: Sets the voltage level to “keep”
the voltage of the previous step.
SP: Linearly increases or decreases
the values from the end of the
previous step to the end of the
current step.
Note
It is available for AC+DC-INT and AC-INT modes only.
DCV
Sets the DC voltage level. There are 3
secondary voltage settings that
determine how the voltage is output.
CT: Sets the voltage level of the step
immediately to DCV values.
KP: Sets the voltage level to “keep”
the voltage of the previous step.
SP: Linearly increases or decreases
the values from the end of the
previous step to the end of the
current step.
Note
It is available for AC+DC-INT and DC-INT modes only.