Parameter
Specification
Hybrid sensors
0.1 – 50 Hz
Velocity output bandwidth
High pass filter output flat to acceleration
0.01 Hz –
spec
*
High pass filter output flat to velocity
spec
– 50 Hz*
Mass position output
DC – 0.1 Hz
Velocity sensitivity
1400 V/m/s
Acceleration sensitivity
2000 V/m/s
2
Velocity sensors
Velocity output bandwidth
spec
– 50 Hz*
Mass position output
DC –
spec
Hz*
Velocity sensitivity
2 × 750 V/m/s
Mass position sensitivity
1000 V/m/s
2
Mass locking and unlocking
manual; optional remote
operation
Mass centring
automatic, MPU controlled
Sensors
3 orthogonal, each 0.180 kg
Lowest spurious resonance
above 140 Hz
Total weight
11 kg
Sensor transducer type
capacitive displacement
Feedback transducer type
magnet/coil
Connector
pressure tight
Temperature range with masses locked
–35 to +75 °C
Operational temperature range
–20 to +65 °C**
Input voltage
10 – 36 V
Current at 12 V DC
75 mA†
Current at 12 V DC during calibration
100 mA†
Current at 12 V DC during centring
330 mA†
Current at 12 V DC during locking and unlocking
490 mA†