Calibrating the 3ESP
[
instrument-id
]
Serial-Nos=
T3X99
VPC=
3.153
,
3.147
,
3.159
G=
1010
,
1007
,
1002
COILCONST=
0.02575
,
0.01778
,
0.01774
CALVPC=
3.161
CALRES=
51000
TYPE=
sensor-type
RESPONSE=
response-code
12.
Click
OK
. The script will return with a graph showing the response of the
sensor in terms of amplitude and phase plots for each component (if
appropriate.)
13.
The accuracy of the results depends on the amount of data you have selected,
and its sample rate. To obtain good-quality results at low frequency, it will save
computation time to use data collected at a lower sample rate; although the
same information is present in higher-rate streams, they also include a large
amount of high-frequency data which may not be relevant to your purposes.
14.
The
bbnoisecal
script automatically performs appropriate averaging to
reduce the effects of aliasing and cultural noise.
4.3.1 Sensor response codes
Sensor
Sensor type code
Units
(V/A)
CMG-5T or 5TD, DC – 100 Hz response
CMG-5_100HZ
A
CMG-40T-1 or 6T-1, 1 s – 100 Hz response
CMG-40_1HZ_50HZ
V
CMG-40_1S_100HZ
V
CMG-40T-1 or 6T-1, 2 s – 100 Hz response
CMG-40_2S_100HZ
V
25