14
12. Packing List
The components of the equipment are listed in the packing
list.
13. Troubleshooting
Faulty
Possible cause
Solution
Non-indication
Non-act.
l. Battery has run down.
2. Battery is fixed upside
down.
3. The LED.C works
continuously too long and
the heat protection system
works.
1. Charge the LED.C.
2.Refix the battery, pay
attention to the ends of the
battery.
3. Stop the operation for
several minutes.
Light intensity
insufficient.
1. The optical fiber is not
inserted till the botton.
2. The optical fiber has
cracked
3. There is resin remained
on the surface of optical
fiber.
4. Low battery.
1. Insert the optical fiber
again correctly.
2. Change the optical fiber.
3. Wipe off the resin.
4 Charge the LED.C.
Show "Er".
1. The battery's voltage is
low.
2. The LED.C is damaged.
1. Charge the LED.C.
2. Contact the local
distributor or us.
Battery lasting
time becomes
shorter.
The battery's capacity
becomes smaller.
Change a new battery.