CURING LIGHT LED.C

USER’S MANUAL

Summary of Contents for CURING LIGHT LED.C

Page 1: ...CURING LIGHT LED C USER S MANUAL ...

Page 2: ...Operation 3 7 Cleaning Disinfection and Sterilization 4 8 Precaution 12 9 Contraindication 13 10 Maintenance 13 11 After service 13 12 Packing List 14 13 Troubleshooting 14 14 Storage and transportation 15 15 Environmental protection 15 16 Manufacturer s right 16 17 Symbol instruction 16 18 Statement 17 19 EMC Declaration of conformity 17 ...

Page 3: ... short time 2 2 This product is used to restore teeth and solidify material for whitening teeth 3 Structure and components LED C dental is composed mainly of high power LED optical fiber and main unit Picture 1 4 Technical specifications 4 1 Power supply Rechargeable Lithium battery Battery model ICR18490 Battery voltage and capacity 1400mAh Input of adapter 100V to 240V 50Hz 60Hz Built in fuse of...

Page 4: ...tion 5 Instruction of recharging 5 1 Load the battery into the compartment of the pedestal charger Be sure not to opposite the electrode 5 2 Connect adapter to the AC100V 240V Machine standby when indicator lighting green Put the battery into the pedestal when indicator turns yellow starts recharging After recharging indicator turns green and shut off the electricity and take out the battery from ...

Page 5: ...er to the end 6 3 To install the light hood on as show in picture 1 6 4 Press the time button to choose the solidification time 4 working time modes are available 5 10 15 20 seconds 6 5 Lightly press the mode key Following three modes are available 6 5 1 Full power mode blue light shine in full power 6 5 2 Ramping mode The power of the blue light turns from weak to stronger and reaches the highest...

Page 6: ... volume recharge it at once 6 11 Please take out the battery and store it carefully when not operating regularly 6 12 The curing light is equipped with over heat protection system It can continuously work 200s For example continuously operate the curing light for 10 times under 20s working mode even the curing light works less than 20s it is counted as a full operation then it will come into over ...

Page 7: ...onsibility for the sterility of products during use only sufficiently validated equipment and product specific procedures are used for cleaning disinfection and sterilization and that the validated parameters are adhered to during every cycle Please also observe the applicable legal requirements in your country as well as the hygiene regulations of the hospital or clinic especially with regard to ...

Page 8: ... should be performed no later than 24 hours after the operation The cleaning can be divided into automated cleaning and manual cleaning Automated cleaning is preferred if conditions permit 7 3 1 Automated cleaning The cleaner is proved to be valid by CE certificationin accordance with ENISO 15883 There should be a flushing connector connected to the inner cavity of the product The cleaning procedu...

Page 9: ...ce with EN ISO 15883 Use high temperature disinfection function The temperature does not exceed 134 and the disinfection under the temperature cannot exceed 20 minutes The disinfection cycle is in accordance with the disinfection cycle in EN ISO 15883 Cleaning and disinfecting steps by using Washer disinfector 1 Carefully place the product into the disinfection basket Fixation of product is needed...

Page 10: ...ture should not exceed 45 C otherwise the protein will solidify and it is difficult to remove c3 The solution used can be pure water distilled water deionized water or multi enzyme solution etc and only freshly prepared solutions can be used c4 During the use of cleaner the concentration and time provided by manufacturer shall be obeyed The used cleaner is neodisher MediZym Dr Weigert d Disinfecti...

Page 11: ...oth the product drying is completed 2 It can also be dried directly in a medical drying cabinet or oven The recommended drying temperature is 80 120 and the time should be 15 40 minutes Notes a The drying of product must be performed in a clean place b The drying temperature should not exceed 138 C c The equipment used should be inspected and maintained regularly 7 6 Inspection and maintenance In ...

Page 12: ...t steam permeability c The packaging environment and related tools must be cleaned regularly to ensure cleanliness and prevent the introduction of contaminants d Avoid contact with parts of different metals when packaging 7 8 Sterilization Use only the following steam sterilization procedures fractional pre vacuum procedure for sterilization and other sterilization procedures are prohibited 1 The ...

Page 13: ... manufacturer assumes no responsibility for the procedures that have not been recommended If you use the sterilization procedures that have not been recommended please adhere to related effective standards and verify the suitability and effectiveness Fractional pre vacuum procedure steam sterilization with repetitive pre vacuum The procedure used here is to perform steam sterilization through thre...

Page 14: ...owel soaked in 75 medical alcohol Repeat the wipe for at least 3 times After each use wipe the surface of the device with a soft cloth soaked in clean water distilled or deionized water or a clean disposable wipe Repeat the wipe for at least 3 times 8 Precaution 8 1 Please recharge the battery at least 4 hours before first time usage 8 2 Connect adapter to the AC100V 240V Machine standby when indi...

Page 15: ...ght works for 40s continously the temperature of the top of optical fiber may reach 56 WARNING Do not modify this equipment without authorization of the manufacturer 9 Contraindication The heart disease patients pregnant women and children should be cautious to use this equipment 10 Maintenance 9 1 Only the optical fiber can be autoclaved under high temperature and pressure 9 2 After operation eac...

Page 16: ...battery 3 Stop the operation for several minutes Light intensity insufficient 1 The optical fiber is not inserted till the botton 2 The optical fiber has cracked 3 There is resin remained on the surface of optical fiber 4 Low battery 1 Insert the optical fiber again correctly 2 Change the optical fiber 3 Wipe off the resin 4 Charge the LED C Show Er 1 The battery s voltage is low 2 The LED C is da...

Page 17: ...s equipment should be handled carefully kept away from shaking point installed or stored at shadowy dry cool and ventilated places 14 2 Don t store it together with articles that are combustible poisonous caustic and explosive 14 3 This equipment should be stored in the environment where the relative humidity is10 93 the atmosphere pressure is 70kPa to 106kPa and the temperature is 20 to 55 14 4 E...

Page 18: ...me differences between blueprint and real equipment take the real equipment as the norm 17 Symbol instruction CE marked product Type B applied part Ordinary equipment Class Ⅱ equipment Date of manufacture Alternating current Manufacturer Recovery Used indoor only Keep dry Screw inside outside Handle with care Sterilizable up to the temperature specified Temperature limitation for storage Humidity ...

Page 19: ... interpretation rights belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO LTD The industrial design inner structure etc have claimed for several patents by WOODPECKER any copy or fake product must take legal responsibilities 19 EMC Declaration of conformity The device has been tested and homologated in accordance with EN 60601 1 2 for EMC This does not guarantee in any way that this device will not...

Page 20: ...18 ...

Page 21: ...19 ...

Page 22: ...20 ...

Page 23: ...21 ...

Page 24: ...ZMN BZ 263 V3 9 2020804 ...