Section 4
OM6622T-A-00
30 January
2018
4-23
4.6.6.
<Edit> MENU (Probe) Fn-2 (Edit MAIN)
The <Edit> menu will either be identical to the standard menu seen in the 6622T
Series bridge, or be as described below dependent on the selection of the
Resistors/Probes mode toggle key in the <Setup> menu.
Mode: RS
:
Display: RS SN:
Update: R0:
Filter: PR SN:
Power: RRate:
A:
══════════╤══════════╤═════════┬──────────
IPTS 68 │ITS 90 │IEC 751 │Probe Set
Function Key <Fn-1> ,<IPTS 68>, allows entry to the IPTS-68 measurement menu
and enables the IPTS-68 coefficients for the probe to be viewed and or edited.
Function Key <Fn-2> ,<ITS 90>, allows entry to the ITS-90 measurement menu and
enables the ITS-90 coefficients for the probe to be viewed and or edited.
Function Key <Fn-3> ,<IEC 751>, allows entry to the IEC-751 measurement menu
and enables the IEC-751 coefficients for the probe to be viewed and or edited.
Function Key <Fn-4> ,<Probe Set>, allows entry to the Probe set menu and enables
the test values for the probe to be viewed and or edited.
4.6.6.1.
<IPTS 68> MENU Fn-1 (Edit MAIN)
A:
B:
b4:
e4:
──────────┬──────────┬─────────┬──────────
Edit │Ok
│ │
Press <Edit> and use arrow keys to move the highlight to the data line to be
changed. Modify data and press <Ok>.
When <Edit> is pressed the PASSWORD CHECK display menu is activated.
Before the selected coefficient can be changed the correct password must be entered.
If an incorrect password is entered, the error message "Invalid Password" will be
displayed and the edit session will be terminated.
Summary of Contents for 6622T Series
Page 2: ......
Page 8: ...OM6622T A 00 30 January 2018 vii...
Page 10: ...OM6622T A 00 30 January 2018 ix...
Page 16: ...Section 1 OM6622T A 00 30 January 2018 1 6...
Page 37: ...Section 3 OM6622T A 00 30 January 2018 3 11 Figure 3 5 Rear Panel Connections PRT SPRT...
Page 80: ...Section 4 OM6622T A 00 30 January 2018 4 42...
Page 92: ...Section 6 OM6622T A 00 30 January 2018 6 6...