UtilityScan DF Quick Start Guide
Page 17
Maintenance
•
Calibrate Survey Wheel (SW): Selecting this option will bring up the survey wheel calibration window. GSSI
recommends calibrating the survey wheel before performing a survey.
1
Select calibration distance using the up/down arrow buttons.
2
Position Antenna on the Start Mark.
3
Click START button
4
Move Antenna to Stop Mark. Take care to ensure that the portion of the antenna which began at the start mark is the
same portion which ends at the stop mark. Press the Stop button.
5
You will see the new calibration value displayed at the bottom of the screen. It will be negative for the
3-wheeled cart and positive for the 4-wheeled cart. Tap on Accept.
•
Configure GPS: Select this option to configure this system with your GPS. See Appendix D for GPS configuration
instructions.
•
Set Time: Select this option to set the system Date and Time.