USER GUIDE
Gas Sensing Solutions Ltd.
Revision 1.2, 29 March 2021
Page | 21 For regular updates, sign up at https://gassensing.co.uk
Copyright © 2021 Gas Sensing Solutions Ltd.
ZERO-POINT SETTING THE SENSOR
There are a several methods available to the user to set the zero point of the sensor. In all cases, the
best zero is obtained when the gas concentration is stable, and the sensor is at a stable temperature.
Note the zero-point settings are not cumulative and only the latest zero-point setting is effective. For
example, there is no benefit in zeroing in nitrogen, and then zeroing in a calibration gas. The sensor
will store only the latest zero point.