The CozIR®-Blink-N takes a number of measurements and filters them to create a single reading.
Each measurement takes 200ms plus the sensor setup time.
For an ‘nPulse’ setting of 16, the time
to report a reading is approximately 16*200ms + 300ms = 3.5s. If the time required by the sensor to
report a valid reading is lower than the minimum cycle time, the software will not accept the value
and the user will be reminded of the minimum value allowable
for the ‘nPulse’ setting.
After the
sensor has reported a valid reading, it is automatically powered down by the software.
To power cycle the sensor manually
, set ‘Measurement Time’ to Manual Measurement.
The sensor
will be power cycled,
and a new reading will be taken when the user clicks ‘Manual Polling –
CO
2
’.
Manually power cycling the sensor faster than the minimum time needed to create a valid reading
will result in erratic sensor performance.
After a manual power cycle, or after each power cycle period, the CO
2
reading will be updated in the
main sensor data display area.