5.2 Deaerating the pump
1. Open deaeration screw by approximately half a
turn.
2. Press and hold down the 100 % key (deaeration
key) until liquid flows continuously without any
bubbles from the deaeration hose.
3. Close deaeration screw.
5.3 Calibrating the pump
The pump is calibrated in the factory for media with a
viscosity similar to water at maximum pump
backpressure (see section
If the pump is operated with a backpressure that
deviates or if dosing a medium whose viscosity
deviates, the pump must be calibrated.
For pumps with FCM control variant, it is not
necessary to calibrate the pump if there is deviating
or fluctuating backpressure as long as the
’AutoFlowAdapt’
function has been enabled
(see section
).
Requirements
•
The hydraulics and electrics of the pump are
connected (see section
•
The pump is integrated into the dosing process
under operating conditions.
•
The dosing head and suction hose are filled with
dosing medium.
•
The pump has been deaerated.
Warning
The deaeration hose must be
connected correctly and inserted into a
suitable tank!
Note
Press the 100 % key and
simultaneously turn the clickwheel
clockwise to increase the duration of
the process to up to 300 seconds. After
setting the seconds, do not press the
key any longer.