Grundfos Dosing monitor Installation And Operating Instructions Manual Download Page 5

5

2.1 Definitions 

Correct dosing stroke:

 A pulse from the dosing 

monitor corresponds to the internal stroke signal 
within acceptable time.

Incorrect dosing stroke: 

There is no pulse from the 

dosing monitor corresponding to the internal stroke 
signal within the acceptable time (the pump is not 
pumping).

2.2 Logic 

If two incorrect dosing strokes are performed, the 
pump will continue operating, but it will change over 
to alarm mode, like in the case of "low level". The red 
indicator light will be on and the alarm output, if any, 
will be activated (variant AR). 

For each correct stroke the counter is reset and the 
alarm output, if any, is deactivated.

2.3 Operation with vent valve

Fig. 2 

It is possible to achieve a self-acting solution for liq-
uids which may cause gas accumulation by using an 
automatic vent valve together with the dosing moni-
tor and an alarm output (control variant AR). The 
alarm output can thus activate the vent valve, which 
automatically vents the gas accumulated in the dos-
ing head. When the dosing head has been vented, 
the pump will start pumping again, the alarm output 
will be deactivated and the vent valve will close.

TM

02

 40

99

 48

01

N

F

NO C

N

F

Reviews: