SLT 5.0 Level and Flow Monitor
Optional PZ12T_F
Maximum Range:
12 ft (3.66 m)
Minimum Range (Deadband):
8"(203.2 mm)
Operating Frequency:
92 KHz
Beam Angle:
8°
Operating Temperature:
-40° to 150° (-40° to 65°C)
Temperature Compensation:
Automatic, continuous
Max. Operating Pressure:
20 psi (1.35 bar)
Sensor Face:
Teflon
Sensor Body:
PVC
Mounting:
¾" NPT
Cable Length:
25 ft. (7.6 m) continuous RG62AU coaxial. Optional 50 ft. (15 m)
continuous
Max. Cable Length:
500 ft. (152 m) RG62AU coaxial (splice)
Hazardous Rating:
CSA rated Intrinsically Safe Class I, Groups C,D, Class II, Groups E,F,G
with optional Intrinsic Safety Barrier.