Page 12
SLT 5.0 Level and Flow Monitor
CALIBRATION
- for Level/Inventory Applications
1. Before starting the calibration determine:
a)
MAX RANGE = ________________ (Maximum
range = distance from Sensor to Zerolevel)
b)
MAX LEVEL = ________________
(Maximum level of product being measured)
c)
MIN RANGE
= ________________
(Distance from sensor to Max Level)
Minimum range = MAX RANGE - MAX LEVEL
(must be at least 8" / 203 mm depending on sensor model)
2. Check the maximum range with the sensor installed by:
a)
When liquid is at zero level press
to view the Range reading in the
Status menu. Use this range measured by the SLT 5.0 as the Max Range
setting.
or
b)
Carefully measure distance from sensor to zero level with a tape measure,
and use this measurement as the Max Range setting.
MINIMUM RANGE:
At CALIBRATION
- Press
to get to Min Range
- Press
to move the cursor under the digits
- Use
or
to set the minimum range
Note
: Min Range must be at least 12" (30.5 cm) for PZ32T sensors, 16" (40.6 cm)
for PZ52T and PZ34 sensors and 8" for PZ15 sensors. For correct tank volume
calculation in Hrt mode Min Range must be the actual distance from the end of
the sensor to the top of the tank.
MAXIMUM RANGE:
Press
to get to Max Range
Enter Max Range determined above.
Summary of Contents for SLT 5.0
Page 2: ...Note This page has been left blank intentionally ...
Page 5: ...Page 5 SLT 5 0 Level and Flow Monitor CONNECTIONS ...
Page 7: ...Page 7 SLT 5 0 Level and Flow Monitor CALIBRATION MENU ...
Page 14: ...Page 14 SLT 5 0 Level and Flow Monitor CHANNEL SETUP ...
Page 20: ...Page 20 SLT 5 0 Level and Flow Monitor SENSOR MOUNTING METHODS ...