From the Data Logging
Store? Yes
prompt the menu will return to
RUN STOP SETUP
. Press
Æ
to
position the cursor under
RUN
and press
È
to ac ti vate the Data Log ger to start. The AVFM-II will
dis play
SESSION NO. x
. Press
È
to re turn to
DATA LOGGING
.
Note: Greyline Logger software cannot accurately calculate totals from 'event' based log files. Use
'trend' logging format if totals must be calculated.
Logging "Sessions"
Each time you select
STOP
in the
DATA LOGGING
menu, the Data Log ger stores the cur rent data in
mem ory as a "
SESSION NO
" au to mat i cally num bered from "
1
" to "
10
". If you re sume log ging by
se lect ing
RUN
, the Data Log ger will re port that a new log ging ses sion is started and ti tled "
SESSION
NO xx
". When you down load the log ger files to your PC us ing Greyline Log ger soft ware, each Ses sion
will open as a sep a rate graph/ta ble ti tled "Greyline Data Log xx".
Important
:
If you
STORE
instrument calibration changes under the
UNITS/MODE
or
CALIBRATION
menus,
STOP
the data logger and select
RUN
again to start a new logging Session with your new calibration
values.