GREISINGER GMH 3610 Operating Manual Download Page 4

H36.0.01.6C-13 

Operating Manual GMH 3610 

page 4 of 12

 

 

7. 

Caution, acid!

 

The electrode contains 

KOH

. KOH can cause severe chemical burns! 

If leaking, avoid contact! 

If there was contact:

 

 

• to skin: flush contacted area with large amounts of water for several minutes. 

 

• to clothing: remove contaminated clothing. 

 

• to eyes: flush with large amounts of water for several minutes,  

   obtain medical treatment. 

After swallowing:

 

 

• give large volumes of water. DO NOT induce vomiting! 

 

• obtain medical treatment. 

Display and control elements 

5.1  Display elements 

 

Display with two sensors connected: 

 

 

 

Main display:

 

possible views: 

 

Oxygen saturation in %

 (% O

2

 

Oxygen concentration 

(mg/l) 

 

 

Secondary display

:    

 

Sensor temperature

 (°C or °F) 

The secondary display can be configured by means 
of the configuration system (p.r.t. 

'Configuration'

 

The desired view can be selected by pressing the 

-key. 

 

Special display elements: 

 

 

Min/Max/Hold

:  

shows if a min., max. or hold value is displayed in 
either the main or the secondary display. 

 

ok arrow

:  

indicates that oxygen and temperature values have 
been stable for a longer period of time 

 

CAL arrow

:  

indicates that an automatic oxygen calibration is car-
ried out

 

 

Warning triangle:  

indicates a low battery, full logger storage, etc. 

 

 

 

 

Display at reboot: 

After switching on the instrument performs a segment test of 2 seconds. 
Afterwards some configurations are displayed sequentially: air pressure and if activated: offset settings. 
(p.r.t Chapter 6 'Configuration') 
 

Tip:  The display of these settings can be aborted by pressing a key after the segment test (keys 2 - 6).

 

 

 

°C

   

2

Logg

AL

mg/l

ok

mmHg

hPa

SAL

hPa

%O

CAL

MAX

MIN HLD

%

°C

   

2

Logg

AL

mg/l

ok

mmHg

hPa

SAL

hPa

%O

CAL

Summary of Contents for GMH 3610

Page 1: ...ndort Greisinger Hans Sachs Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Hand held measuring device for dissolved oxygen and temperature as of version V2 1 Operating...

Page 2: ...AND SYSTEM MESSAGES DURING OXYGEN CALIBRATION 10 12 THE SERIAL INTERFACE 11 12 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 11 13 SPECIFICATION 12 14 DISPOSAL INSTRUCTION 12 1 Designated Use...

Page 3: ...cordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given...

Page 4: ...be configured by means of the configuration system p r t Configuration The desired view can be selected by pressing the key Special display elements Min Max Hold shows if a min max or hold value is d...

Page 5: ...ec start oxygen calibration Set Menu press Set shortly changing of the main display press Menu for 2 sec configuration will activated Store Quit Measurement Hold current measuring value HLD in display...

Page 6: ...us F All temperature values in degrees Fahrenheit Power off Selection of Power Off Delay 1 120 Power off delay in minutes Device will be automatically switched off as soon as this time has elapsed if...

Page 7: ...rode sup plies a current No oxygen means no current either The oxygen measurements use up both the silver cathode as well as the lead anode The electrode is subject to ageing Therefore we recommend to...

Page 8: ...to calibrate oxygen electrode The temperatures of the electrode and of the liquid to be measured have to be identical if necessary wait till temperatures match The Electrode has to be submerged at lea...

Page 9: ...the diaphragm get wet or be immersed in water The bottle neck opening should only be slightly larger than the electrode diameter make sure to avoid over pressure in the container Note A correctly cal...

Page 10: ...probe Values below measuring range Are there any values below the measuring range specified measuring value too low Sensor cable defective replace probe System fault switch on again if fault continue...

Page 11: ...pment package including a universally applicable Windows functions library GMH3000 DLL with documentation can be used by all es tablished programming languages suitable for Windows XP Windows Vista Wi...

Page 12: ...r GRS3105 see accessories Power supply 9V battery type IEC 6F22 included or additional d c connector internal pin 1 9 mm for external 10 5 12V direct voltage supply suitable power supply GNG10 3000 Po...

Reviews: