GREISINGER GMH 3610 Operating Manual Download Page 3

H36.0.01.6C-13 

Operating Manual GMH 3610 

page 3 of 12

 

 

Operating and Maintenance Advice 

a) 

First Usage 
If the electrode was disconnected from the device, pull off the protection flask and expose the electrode at least 2 - 3 hours to 
the air before the first calibration or measuring.

 

a)  When to replace battery: 

If 

 and ´bAt´ is shown in the lower display the battery has been used up and needs to be replaced. The device 

will, however, operate correctly for a certain time. 
If ´bAt´ is shown in the upper display the voltage is too low to operate the device; the battery has been completely 
used up. 

Please note:  The battery has to be taken out, when storing device above 50°C. 

We recommend to take out battery if device is not used for a longer period of time. 

b)  Treat device and sensor carefully and use it only in accordance with its specifications (do not throw, hit against 

etc.). Protect plug and socket from soiling. 

c)  Make sure to use only sensors that are suitable for the device. Unsuitable measuring probes may lead to the de-

struction of the measuring device and the measuring probes. 

d)  When connecting the electrode the connector may not lock to the jack correctly. In such a case hold the connector 

not at the case but at the buckling protection of the cable during the plug in. 
Don't connect electrode canted! If plug is entered correctly, it will slide in smoothly. 
To disconnect sensor do not pull at the cable but at the plug 

e)  Mains operation: 

When using a power supply device please note that operating voltage has to be 10.5 to 12 V DC. 
Do not apply overvoltage!! Cheap 12V-power supply devices often have excessive no-load voltage. Therefore, 
we recommend using regulated voltage power supply devices. Using the power supply device GNG10/3000 en-
sures trouble-free operation.  
Prior to connecting the power supply device to the mains supply make sure that the operating voltage stated at 
the power supply device is identical to the mains voltage. 

Safety Requirements 

This device has been designed and tested in accordance with the safety regulations for electronic devices. 
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and 
special safety advises given in this manual will be adhered to when using the device. 

1.  Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any 

other climatic conditions than those stated under "Specifications". 

2.  If the device is transported from a cold to a warm environment condensation may cause in a failure of the func-

tion. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a 
new start-up. 

3.  If device is to be connected to other devices (e.g. via serial interface) the circuitry has to be designed most care-

fully. Internal connection in third party devices (e.g. connection GND with earth) may result in not-permissible 
voltages impairing or destroying the device or another device connected. 

 

Warning: 

If device is operated with a defective mains power supply (short circuit from mains voltage to output 

voltage) this may result in hazardous voltages at the device (e.g. sensor socket, serial interface). 

4.  If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be 

marked accordingly to avoid re-starting. 

 

Operator safety may be a risk if: 

 

 - there is visible damage to the device 

 

 - the device is not working as specified 

 

 - the device has been stored under unsuitable conditions for a longer time. 

 

In case of doubt, please return device to manufacturer for repair or maintenance. 

5. 

Warning: 

Do not use these products as safety or emergency stop devices or in any other application where failure 

of the product could result in personal injury or material damage. 
Failure to comply with these instructions could result in death or serious injury and material damage. 

6.  This device only serves as monitoring of systems important for the customer. 

It must not be used instead of compulsory approval monitoring devices and it is not planned for that pur-
pose. If this device is used for the monitoring of such systems on its own, the manufacturer will not as-
sume liability for damages whatsoever. 

Summary of Contents for GMH 3610

Page 1: ...ndort Greisinger Hans Sachs Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Hand held measuring device for dissolved oxygen and temperature as of version V2 1 Operating...

Page 2: ...AND SYSTEM MESSAGES DURING OXYGEN CALIBRATION 10 12 THE SERIAL INTERFACE 11 12 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 11 13 SPECIFICATION 12 14 DISPOSAL INSTRUCTION 12 1 Designated Use...

Page 3: ...cordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given...

Page 4: ...be configured by means of the configuration system p r t Configuration The desired view can be selected by pressing the key Special display elements Min Max Hold shows if a min max or hold value is d...

Page 5: ...ec start oxygen calibration Set Menu press Set shortly changing of the main display press Menu for 2 sec configuration will activated Store Quit Measurement Hold current measuring value HLD in display...

Page 6: ...us F All temperature values in degrees Fahrenheit Power off Selection of Power Off Delay 1 120 Power off delay in minutes Device will be automatically switched off as soon as this time has elapsed if...

Page 7: ...rode sup plies a current No oxygen means no current either The oxygen measurements use up both the silver cathode as well as the lead anode The electrode is subject to ageing Therefore we recommend to...

Page 8: ...to calibrate oxygen electrode The temperatures of the electrode and of the liquid to be measured have to be identical if necessary wait till temperatures match The Electrode has to be submerged at lea...

Page 9: ...the diaphragm get wet or be immersed in water The bottle neck opening should only be slightly larger than the electrode diameter make sure to avoid over pressure in the container Note A correctly cal...

Page 10: ...probe Values below measuring range Are there any values below the measuring range specified measuring value too low Sensor cable defective replace probe System fault switch on again if fault continue...

Page 11: ...pment package including a universally applicable Windows functions library GMH3000 DLL with documentation can be used by all es tablished programming languages suitable for Windows XP Windows Vista Wi...

Page 12: ...r GRS3105 see accessories Power supply 9V battery type IEC 6F22 included or additional d c connector internal pin 1 9 mm for external 10 5 12V direct voltage supply suitable power supply GNG10 3000 Po...

Reviews: