13  Specification 

Measuring range:  

 

Oxygen concentration  

0.0...25.0 mg/l  

Oxygen saturation  

0...300 %  

Electrode temperature  

0.0

…50.0 °C

 

 

 

 

Accuracy in the device  

(at nominal temperature) 

Oxygen meas. ranges  

±1.5% ±0.2mg/l 

Electrode temperature  

±0.1°C ±1digit 

Atmospheric pressure compensation: 

setable: 500 .. 2000 hPa abs.

 

Nominal temperature: 

 

20 °C 

Working temperature: 

 

0 to +50 °C 

Relative humidity:

  

0 to 95 %RH (non-condensing) 

Storage temperature: 

 

-20 to +70 °C (electrode: 0 to +60°C)

 

Sensor connection:  

6-pin Mini-DIN-socket 

Electrode  

oxygen electrode (active diaphragm type) with integrated NTC resistor 

Response time:  

95% in 10 sec., depending on temperature 

Operation life:  

guaranteed 12 month (assuming appropriate usage and on proper maintenance) 

Operating pressure:  

max. 3 bar 

Mounting diameter:  

12.0 ±0.2 mm (also suitable for 1/2"glanding) 

Length:  

approx. 220 mm 

Mounting length:  

approx. 110 mm 

Weight:  

approx. 180 g 

Cable length:  

4 m 

Working temperature:  

0 to 40 °C 

Display: 

2 four digit LCDs (12.4 mm high and 7 mm high) for temperature, and for min./ max 
values, hold function, etc. as well as additional pointing arrows. 

Pushbuttons:  

6 membrane keys altogether for on/off switch, selection of thermoelements, min. and 
max. value memory, hold-function etc. 

Interface:  

serial interface (3.5 mm jack), serial interface can be directly connected to USB or 
RS232 interface of a PC via interface adapter USB3100, GRS3100 or GRS3105  
(see accessories). 

Power supply:  

9V-battery, type IEC 6F22 (included) or additional d.c. connector (internal pin Ø 1.9 mm) 
for external 10.5-12V direct voltage supply. 

 

(suitable power supply: GNG10/3000) 

   Power consumption:  

approx. 1.5 mA, during audio alarm approx. 2 mA

 

Automatic-off-function:  

Device will be automatically switched off if no key is pressed/no interface communica-
tion takes place for the time of the power-off delay. The power-off delay can be set to 
values between 1and 120 min.; it can be completely deactivated. 

Min-/max-value memory:  

Both the max. and the min. value will be memorized for each measurement taken. 

Hold-function:  

Press button to memorize current measuring values. 

Housing dimensions: 

142 x 71 x 26 mm (L x W x D)  
impact-resistant ABS plastic housing, membrane keyboard, transparent panel. Front 
side IP65, integrated pop-up clip for table top or suspended use. 

Weight: 

approx. 155 g (device incl. Battery) 

EMC:  

The device corresponds to the essential protection ratings established in the Regula-
tions of the Council for the Approximation of Legislation for the member countries re-
garding electromagnetic compatibility (2004/108/EG) 

 

Additional fault: <1% 

14  Disposal instruction: 

Batteries must not be disposed in the regular domestic waste but at the designated collecting points.  
The device must not be disposed in the unsorted municipal waste!  
The electrode contains lead and caustic electrolyte. Dispose as special waste. 
Send the device and the electrode directly to us (sufficiently stamped), if it should be disposed. We 
will dispose the device and the electrode appropriate and environmentally sound. 

 

_

+

Summary of Contents for GMH 3610

Page 1: ...ndort Greisinger Hans Sachs Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Hand held measuring device for dissolved oxygen and temperature as of version V2 1 Operating...

Page 2: ...AND SYSTEM MESSAGES DURING OXYGEN CALIBRATION 10 12 THE SERIAL INTERFACE 11 12 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 11 13 SPECIFICATION 12 14 DISPOSAL INSTRUCTION 12 1 Designated Use...

Page 3: ...cordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given...

Page 4: ...be configured by means of the configuration system p r t Configuration The desired view can be selected by pressing the key Special display elements Min Max Hold shows if a min max or hold value is d...

Page 5: ...ec start oxygen calibration Set Menu press Set shortly changing of the main display press Menu for 2 sec configuration will activated Store Quit Measurement Hold current measuring value HLD in display...

Page 6: ...us F All temperature values in degrees Fahrenheit Power off Selection of Power Off Delay 1 120 Power off delay in minutes Device will be automatically switched off as soon as this time has elapsed if...

Page 7: ...rode sup plies a current No oxygen means no current either The oxygen measurements use up both the silver cathode as well as the lead anode The electrode is subject to ageing Therefore we recommend to...

Page 8: ...to calibrate oxygen electrode The temperatures of the electrode and of the liquid to be measured have to be identical if necessary wait till temperatures match The Electrode has to be submerged at lea...

Page 9: ...the diaphragm get wet or be immersed in water The bottle neck opening should only be slightly larger than the electrode diameter make sure to avoid over pressure in the container Note A correctly cal...

Page 10: ...probe Values below measuring range Are there any values below the measuring range specified measuring value too low Sensor cable defective replace probe System fault switch on again if fault continue...

Page 11: ...pment package including a universally applicable Windows functions library GMH3000 DLL with documentation can be used by all es tablished programming languages suitable for Windows XP Windows Vista Wi...

Page 12: ...r GRS3105 see accessories Power supply 9V battery type IEC 6F22 included or additional d c connector internal pin 1 9 mm for external 10 5 12V direct voltage supply suitable power supply GNG10 3000 Po...

Reviews: