GREISINGER GMH 3430 Operating Manual Download Page 8

H34.0.01.6C-11 

Operating Manual GMH 3430 

page 8 of 10

 

 

12  The Conductivity Measuring Cell

 

 

During the measurement, the conductivity measuring cell must be dipped into so far, that at least 30mm beginning 
from the top of the measuring cell, is located in the medium. 
The maximum immersion depth for continuous operation should be 110mm. 

The measuring cell can either be stored dry or in water. After dry storage wetting time will be prolonged slightly. If 
changing over from one liquid to another with conductivities varying widely make sure to properly rinse and shake 
dry measuring cell. 

 

  Measuring cell must never come into contact with water-repellent materials such as oil or silicone. 

If conductivity measured is much higher or lower than expected this may be due to the electrode being soiled with 
nonconducting or conducting foreign materials. Measuring cell has to be cleaned with a watery soap solution. 

When measuring media with low conductivities the electrode has to be moved sufficiently. 

13  Error And System Messages  

Error or system 
messages: 

Description / Reason: 

Remedy: 

 

Low battery voltage  
device will only continue operation for 
a short time 

replace battery 

 

Low battery voltage 

replace battery 

- If mains operation: wrong voltage 

replace power supply, if fault continues to exist: 
device damaged 

No display

 

or 

mazy characters

 

- Battery voltage too low 

replace battery 

- If mains op.: power supply defective 

or wrong voltage/polarity 

check/replace mains supply 

- System error 

disconnect battery or power supply, wait for a 
short time, re-connect 

- Device defective 

return to manufacturer for repair 

 

Values exceeding measuring range 
 

Check: are there any values exceeding the 
measuring range specified? 
-> meas. device not suitable 

Electrode/sensor/cable defective

 

-> replace electrode/probe

 

 

Values below measuring range 

check: are there any values below the measur-
ing range specified? 
-> meas. device not suitable 

Electrode/sensor/cable defective

 

-> replace electrode/probe

 

 

System fault 

Disconnect battery, wait for 10 sec., switch on 
again: if fault continues to exist, device is dam-
aged  
-> return to manufacturer for repair  

 

Value could not be calculated 

A measuring variable device required for calcu-
lation is faulty (overflow/underflow)  

 

 

30 mm

110 mm

Summary of Contents for GMH 3430

Page 1: ...electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Measuring device for conductivity temperature as of version V2 1 Operating Manual GMH 343...

Page 2: ...SURING CELL 8 13 ERROR AND SYSTEM MESSAGES 8 14 THE SERIAL INTERFACE 9 14 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 9 15 SPECIFICATION 10 16 DISPOSAL INSTRUCTION 10 1 Designated Use The de...

Page 3: ...entical to the mains voltage 4 Safety Requirements This device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reli...

Page 4: ...Configuration to enter values or change settings Range for conductivity measuring only press for 2 sec change between automatic and manual measuring range selection press shortly at manual measuring...

Page 5: ...n by pressing the key for a short time In case you have already been working in the highest resolution range pressing of the key will bring you down to the lowest measuring range resolution The corres...

Page 6: ...C t Lin Setting of Temperature Coefficient only when t Cor Lin 0 300 3 000 Temperature compensation coefficient in K t rEF Selection of Reference Temperature only when t Cor nLF or Lin 20 C 68 F refe...

Page 7: ...hannel 3 23 Using the interface converter GRS3105 it is possible to connect several devices to a single interface As a precondi tion the base addresses of all devices must not be identical In case sev...

Page 8: ...tem Messages Error or system messages Description Reason Remedy Low battery voltage device will only continue operation for a short time replace battery Low battery voltage replace battery If mains op...

Page 9: ...a Programming examples Visual Basic 4 0 Delphi 1 0 Testpoint If you want to use the interface functions make sure to switch off the Auto Range function 14 1 The following interface functions will be s...

Page 10: ...10 5 12V direct voltage supply suitable power supply GNG10 3000 Power consumption approx 3 5 mA Low battery warning and bAt Display 2 four digit LCD displays 12 4mm and 7mm high for meas values min an...

Reviews: