GREISINGER GMH 3430 Operating Manual Download Page 4

H34.0.01.6C-11 

Operating Manual GMH 3430 

page 4 of 10

 

 

Display and control elements 

5.1  Display elements 

       

 

 

Main display

conductivity, resistivity, filtrate dry resi-
due (TDS), salinity or user prompt 

 

MAX/MIN/HLD:  

display elements to show mini-
mum/maximum/ 
memorized measuring value

 

 

Display of 

measuring units 

for main display 

 

Warning signal: 

indicates low battery 

 

Units 

for configuration settings 

 

Lin/nLF-arrow:

  indicates set temperature compensation  

 

Secondary

 

measuring value temperature or 

Display: 

  

configuration settings 

 

Measuring value units 

for temperature 

5.2  Pushbuttons 

 

 

On/off key 

 

 

min/max when taking measurements: 

press shortly: 

indication of min. or max. value meas-
ured so far as well as corresponding 
temperature  

press for 2 sec.:  the corresponding value is deletes 

Configuration

to enter values or change settings 

 

Range

: (for conductivity measuring only) 

press for 2 sec.:  change between automatic and manual 

measuring range selection 

press shortly (at manual measuring range selection): 

change next higher measuring range or 
change from highest to smallest meas-
uring range 

 

Set/Menu

press shortly 

(Set)

: change-over between measuring 

units: conductivity, resistivity of fluid, 
filtrate dry residue (TDS) and salinity 

press for 2 sec. (Menu): configuration will be activated 

 

Store/Quit

measurement:  holding of current meas. value ('HLD' in 

display)  

Set/Menu: 

acknowledge setting, return to measuring  

5.3  Connections 

 

 

 

Interface

: Connection for electrically isolated interface adaptor 

(

accessories: 

GRS 3100, GRS3105 o. USB3100)

 

Electrode

:  cable gland for fixed conductivity electrode  

 

The mains socket is located at the left side of the instrument.

 

 

MAX DIF

MIN HLD

%

°C

°F

   S

Logg

%/K

AL

mS

nLF

SAL

mg/l

1/cm

Lin

kOhm

1

2

3

4

5

6

ON

OFF

max

Range

Store

min

Set

Menu

Quit

Summary of Contents for GMH 3430

Page 1: ...electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Measuring device for conductivity temperature as of version V2 1 Operating Manual GMH 343...

Page 2: ...SURING CELL 8 13 ERROR AND SYSTEM MESSAGES 8 14 THE SERIAL INTERFACE 9 14 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 9 15 SPECIFICATION 10 16 DISPOSAL INSTRUCTION 10 1 Designated Use The de...

Page 3: ...entical to the mains voltage 4 Safety Requirements This device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reli...

Page 4: ...Configuration to enter values or change settings Range for conductivity measuring only press for 2 sec change between automatic and manual measuring range selection press shortly at manual measuring...

Page 5: ...n by pressing the key for a short time In case you have already been working in the highest resolution range pressing of the key will bring you down to the lowest measuring range resolution The corres...

Page 6: ...C t Lin Setting of Temperature Coefficient only when t Cor Lin 0 300 3 000 Temperature compensation coefficient in K t rEF Selection of Reference Temperature only when t Cor nLF or Lin 20 C 68 F refe...

Page 7: ...hannel 3 23 Using the interface converter GRS3105 it is possible to connect several devices to a single interface As a precondi tion the base addresses of all devices must not be identical In case sev...

Page 8: ...tem Messages Error or system messages Description Reason Remedy Low battery voltage device will only continue operation for a short time replace battery Low battery voltage replace battery If mains op...

Page 9: ...a Programming examples Visual Basic 4 0 Delphi 1 0 Testpoint If you want to use the interface functions make sure to switch off the Auto Range function 14 1 The following interface functions will be s...

Page 10: ...10 5 12V direct voltage supply suitable power supply GNG10 3000 Power consumption approx 3 5 mA Low battery warning and bAt Display 2 four digit LCD displays 12 4mm and 7mm high for meas values min an...

Reviews: