H34.0.01.6C-11
Operating Manual GMH 3430
page 7 of 10
’Unit’:
Selection of Temperature Unit
°C /°F
°C
All temperature values in degrees Celsius
°F
:
All temperature values in degrees Fahrenheit
’Offset’:
Selection of Temperature Offset
-2.0°C ... 2.0°C
or
-3.6°F ... 3.6°F
The zero point of the temperature measurement will be dis-
placed by this value:
temperature displayed = temperature measured - offset
The offset is used to compensate for deviations.
off
:
Zero displacement has been deactivated (=0.0°)
’Power.off’:
Selection of Power-off Delay
1...120
:
Power-off delay in minutes. The instrument will be automatically
switched off as soon as this time has elapsed if no key is pressed/no
interface communication takes place.
off:
automatic power-off function deactivated (continuous operation, e.g. in
case of mains operation)
’
Address
’:
Selection of Base Address
01, 11, 21, ..., 91
:
Base address for interface communication.
Channel 1 will be addressed by the base address set, chan-
nels 2 and 3 will have the following addresses.
(Example: base address 21 - channel 1 = 21, channel 2 = 22, channel 3 = 23)
Using the interface converter GRS3105 it is possible to connect several devices to a single interface. As a precondi-
tion the base addresses of all devices must not be identical. In case several devices will be connected via one inter-
face make sure to configurate the base addresses accordingly.
11 Minimum and Maximum Values
The min. or max. value of the current measuring unit will be shown in the upper line of the
display after the
- or
-keys have been pressed for a short time.
The lower display line will simultaneously show the temperature at which the max./min.
value for conductivity or resistivity, filtrate dry residue, salinity has occurred.
To switch over between the max./min. values of the measuring units press
for a short
time.
Max./min. values for conductivity, filtrate dry residue and salinity will be recorded even
though the measuring unit is not shown in the display at the moment.
To delete max./min. values press
-key or
-key for approx. 2 seconds.
°C
S
Logg
%/K
AL
mS
nLF
SAL
mg/l
1/cm
Lin
kOhm
max
min
2
5
°C
S
Logg
%/K
AL
mS
nLF
SAL
mg/l
1/cm
Lin
kOhm
max
min
2
5
S
Logg
%/K
AL
mS
nLF
SAL
mg/l
1/cm
Lin
kOhm
max
min
2
5
S
Logg
%/K
AL
mS
nLF
SAL
mg/l
1/cm
Lin
kOhm
max
min
2
5
MAX
°C
S
Logg
%/K
AL
mS
nLF
SAL
mg/l
1/cm
Lin
kOhm