GREISINGER GMH 3430 Operating Manual Download Page 1

 

H34.0.01.6C-11 

GREISINGER

 

electronic GmbH

 

D - 93128 Regenstauf, Hans-Sachs-Straße 26 

 +49 (0) 9402 / 9383-0  

 +49 (0) 9402 / 9383-33   

 

[email protected]

 

 

Measuring device for   conductivity 

 

 

temperature 

as of version V2.1 

Operating Manual 

GMH 3430

 

 

 

WEEE-Reg.-Nr. DE93889386 

Summary of Contents for GMH 3430

Page 1: ...electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Measuring device for conductivity temperature as of version V2 1 Operating Manual GMH 343...

Page 2: ...SURING CELL 8 13 ERROR AND SYSTEM MESSAGES 8 14 THE SERIAL INTERFACE 9 14 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 9 15 SPECIFICATION 10 16 DISPOSAL INSTRUCTION 10 1 Designated Use The de...

Page 3: ...entical to the mains voltage 4 Safety Requirements This device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reli...

Page 4: ...Configuration to enter values or change settings Range for conductivity measuring only press for 2 sec change between automatic and manual measuring range selection press shortly at manual measuring...

Page 5: ...n by pressing the key for a short time In case you have already been working in the highest resolution range pressing of the key will bring you down to the lowest measuring range resolution The corres...

Page 6: ...C t Lin Setting of Temperature Coefficient only when t Cor Lin 0 300 3 000 Temperature compensation coefficient in K t rEF Selection of Reference Temperature only when t Cor nLF or Lin 20 C 68 F refe...

Page 7: ...hannel 3 23 Using the interface converter GRS3105 it is possible to connect several devices to a single interface As a precondi tion the base addresses of all devices must not be identical In case sev...

Page 8: ...tem Messages Error or system messages Description Reason Remedy Low battery voltage device will only continue operation for a short time replace battery Low battery voltage replace battery If mains op...

Page 9: ...a Programming examples Visual Basic 4 0 Delphi 1 0 Testpoint If you want to use the interface functions make sure to switch off the Auto Range function 14 1 The following interface functions will be s...

Page 10: ...10 5 12V direct voltage supply suitable power supply GNG10 3000 Power consumption approx 3 5 mA Low battery warning and bAt Display 2 four digit LCD displays 12 4mm and 7mm high for meas values min an...