User’s Manual GMH3151

page 8 of 12

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

_______________________________________________________________________________________________________________________

GREISINGER electronic GmbH

3.2  „Func-CYCL“: Automatic Recording With Selectable Logger-Cycle-Time

The Logger-Cycle-Time is settable (p.r.t. Configuration). For example „CYCL“ = 60: A measuring is recorded
after each 60 seconds.
When the slow measurement "rAtE-Slo" is chosen, additionally a low power function is available: „Lo.Po“.
If „Lo.Po“ is on, the device only will take a measurement at the point of time of the recording. In between the
recordings the measuring shuts down. This decreases the power consumption enormously and therefore is
recommended e.g. for long time recordings where no mains adapter is available.

Max. number of measurings: 10000
Cycle time:

1...3600 seconds (=1h), selectable in the configuration

A measuring contains:

!

  slow measuring rate (rAtE Slo):

- current measuring value at the time of recording
-  min peak, max peak since the last recording

!

  fast measuring rates (rAtE FASt,P.dEt):

- arithmetic mean value since the last recording
- min peak, max peak since the last recording

Starting a recording:

By pressing "Store" (key 6) for 2 seconds the logger operation will be called. The display will show: 

By pressing "Store" again the recording will be initiated.

After that the display shows ‘St.XXXX‘ for a short time whenever a measuring is recorded. XXXX is the
number of the measuring 1... 10000.

If the logger memory is full, the display will show: 

 The recording automatically will be stopped.

If Low-Power-Logger-Function „Lo.Po = on“ the device switches itself off as soon as the memory gets filled.

Stopping the recording manually:

By pressing "Store" (key 6) the recording can be stopped manually. Then the following choice appears:

 

The selection can be made by 

5

 (key 2) and 

6

 (key 5). "Quit" (key 6) enters the choice.

Note: 

If you try to switch off the instrument in the cyclic recording operation You will be asked
once again if the recording is to be stopped.
The device can only be switched off after the recording has been stopped!
The Auto-Power-Off-function is deactivated during recording!

Clear Recordings:

By pressing "Store" (key 6) for 2 seconds the logger operation will be called.

The display will show: 

 By pressing the keys 

5

 (key 2) or 

6

 (key 5) the display will change to 

.

When „Store“ is pressed, the choice for clearing the logger memory will be displayed:

Clear all recordings

Clear the last
recording sequence

Clear nothing
(cancel menu)

The selection can be made by 

5

 (key 2) and 

6

 (key 5). "Quit" (key 6) enters the choice.

Stop the
recording

Do not stop the
recording

Summary of Contents for GMH 3151

Page 1: ...ng 5 2 2 MEASURING OF WATER LEVEL DISPLAY UNIT M 5 2 3 SEA LEVEL CORRECTION FOR ABSOLUTE PRESSURE SENSORS 5 2 4 AVERAGING FUNCTION 5 2 5 ZERO DISPLACEMENT SENSOR OFFS 6 2 6 SCALE CORRECTION SENSOR SCA...

Page 2: ...ions for a longer period of time In case of doubt please return device to manufacturer for repair or maintenance 6 Warning Do not use these product as safety or emergency stop device or in any other a...

Page 3: ...3 Set Menu Calling of configuration Tare Function By pressing Tara key 3 the displays will be set to 0 All measurings from then on will be displayed relatively to the set tare values When the tare fun...

Page 4: ...Base address of interface if Out SEr dAC 0 eg 5 00 5 00 mbar Enter desired value at which the analogue output potential should be 0V if Out dAC SEt ConF dAC 1 eg 5 00 5 00 mbar Enter desired value at...

Page 5: ...press in contact to ambient air by means of a tube no water contact and bring the sensor with its open press connection for higher pressure to water depth to be measured display and is compensated fo...

Page 6: ...ces To get access to each device the base addresses of the devices have to be different For example choose 01 for the first 11 for the second device and so on See also chapter 4 2 8 2 Analogue Output...

Page 7: ...key 6 is pressed a measurement and its time stamp will be recorded The recorded data can be viewed either in the display when calling the configuration an additional menu REAd LoGG is displayed see b...

Page 8: ...the logger operation will be called The display will show By pressing Store again the recording will be initiated After that the display shows St XXXX for a short time whenever a measuring is recorded...

Page 9: ...he interface activate the alarm function of the GMH and the relays output is operating The device has 3 channels Channel 1 current measuring value base address Channel 2 min peak p r t chapter 3 Chann...

Page 10: ...E For measurements of over under or absolute pressure screw sensor to G1 4 pressure terminal or plug plastic tube to a suitable adapter 6 Error And System Messages Display Meaning What to do Low batte...

Page 11: ...he min and max rails set Alarming is done by integrated horn display and interface Real time clock Integrated clock with date and year Logger 2 Functions individual value logger Func Stor and cyclic l...

Reviews: