GREISINGER GLF-100 Operating Manual Download Page 4

H66.0.01.6C-01

Operating Manual GLF100

page 4 of 6

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

3  Configuration of the instrument

To configure the instruments according to Your needs proceed like follows::

  Switch off instrument.

 Switch 

on

and press 

mode-key during the segment test

 (                   ), until the display shows the first parameter ‚P_oF‘

  If a parameter should be edited, press key up or down ( 

 ), the setting of the parameter will be shown and

can be changed via up/down keys and receipt the changed parameter with 

.

  Jump to the next parameter with 

parameter

value

meaning

key

 

keys

 

Auto Power-Off

Auto Power-Off

  in minutes. If there won’t be pressed any key and no interface

communication takes place for the time of the power off time setting, the device will be
switched off automatically to save battery power (adjustable range 1 to 120 min)
automatic power-off is deactivated (continuous operating)

Unit of display
Auto
µS/cm+mS/cm

Measuring of conductivity with auto-range (factory setting)

100.0 mS/cm

Measuring of conductivity, range 0.0 ... 100.0 mS/cm (no auto-range)

20.00 mS/cm

Measuring of conductivity, range 0.00 ... 20.00 mS/cm (no auto-range)

2000 µS/cm

Measuring of conductivity, range 0 ... 2000 µS/cm (no auto-range)

2000 mg/l

Measuring of TDS, 0 ... 2000 mg/l

50.0 SAL

Measuring of Salinity, 0.0 .. 50.0 g/kg

Unit of temperature display
°C

temperature is displayed in °C (factory setting)

°F

temperature is displayed in °F

temperature compensation (only for µS/cm and mS/cm, p.r.t. note below)
oFF

No temperature compensation of conductivity measuring

nLF

non-linear temperature compensation for natural water acc. to EN27888 (DIN
38404) to measure ground water, surface water, drinking water and purest water
(factory setting)

Reference temperature of temp. compensation (only for t.Cor <> oFF)
25°C/77°F

Reference temperature 25°C / 77°F(factory setting)

20°C/68°F

Reference temperature 20°C / 68°F

Setting of TDS-factor
0.40...1.00

Setting of cell constant for adjusting of the measurement

Reset settings to Ex-Works Settings

Settings are kept

Settings of configuration and adjustment menu are reset to ex-works settings

Pressing 

 again stores the settings, the instruments restarts (segment test)

Please note:

If there is no key pressed within the menu mode within 2 minutes, the configuration will be
cancelled, the entered settings are lost!

Note for temperature compensation: To measure the TDS value always use non-linear temperature

compensation for natural water (reference temperature = 25°C).
To measure salinity the instrument automatically switches over to the non-linear temperature
compensation acc. to IOT (reference temperature = 15°C)..

Summary of Contents for GLF-100

Page 1: ...UCTIVITY MEASURING 3 2 1 THE CONDUCTIVITY MEASURING CELL 3 2 2 MEASURING HINTS 3 2 3 TEMPERATURE COMPENSATION 3 2 4 MIN MAX VALUE MEMORY 3 2 5 HOLD FUNCTION 3 3 CONFIGURATION OF THE INSTRUMENT 4 4 ADJUSTING OF THE INSTRUMENT 5 5 SYSTEM MESSAGES 5 6 INSPECTION OF THE ACCURACY ADJUSTMENT SERVICES 5 7 SPECIFICATION 6 precise measurements of conductivity TDS total dissolved solids and Salinity Applica...

Page 2: ...ergency stop device or in any other application where failure of the product could result in personal injury or material damage Failure to comply with these instructions could result in death or serious injury and material damage 1 2 Operating And Maintenance If the symbol LO BAT is displayed at the left side of display the battery is weak measuring can be continued for a short period If bAt is di...

Page 3: ...hem useless very fast Measuring procedure Especially when measuring low conductivity Before immersion to the measuring solution rinse the electrode with deionised water dry with a paper towel and shake remaining water out The measuring is speeded up considerably if the electrode is immersed and pulled out the solution several times Especially when measuring low conductivity the electrode needs suf...

Page 4: ...ng of TDS 0 2000 mg l 50 0 SAL Measuring of Salinity 0 0 50 0 g kg Unit of temperature display C temperature is displayed in C factory setting F temperature is displayed in F temperature compensation only for µS cm and mS cm p r t note below oFF No temperature compensation of conductivity measuring nLF non linear temperature compensation for natural water acc to EN27888 DIN 38404 to measure ground...

Page 5: ...emperature 2 0 2 0 C 3 6 3 6 F Offset setting of temperature measurement see below Scale of temperature 5 00 5 00 Scale adjustment of temperature measurement Display measured value offset 1 scale adjustment 100 Pressing again stores the settings the instruments restarts segment test Please note If there is no key pressed within the menu mode within 60 seconds the adjusting will be cancelled the en...

Page 6: ... 100 C Relative humidity instrument 0 95 r F non condensing Storage temperature 5 50 C Housing Dimension 110 x 67 x 30 mm L x B x D impact resistant ABS membrane keyboard transparent panel Front side IP65 Weight approx 155 g Power Supply 9V Battery type IEC 6F22 included Power Consumption 1 5mA Display ca 11 mm high 4 digit LCD display with additional segments Pushbuttons 3 membrane keys for on of...

Reviews: