GREISINGER GLF-100 Operating Manual Download Page 2

H66.0.01.6C-01

Operating Manual GLF100

page 2 of 6

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

1 In General

1.1 Safety 

Instructions

This device has been designed and tested in accordance to the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and
special safety advises given in this manual will be adhered to when using it.
1.  Trouble-free operation and reliability of the device can only be guaranteed if it is not subjected to any other climatic

conditions than those stated under “Specification”.

2.  Transporting the device from a cold to a warm environment condensation may result in a failure of the function. In

such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up.

3.  The circuitry has to be designed most carefully if the device should be connected to other devices. Internal

connection in third party devices (e.g. connection GND and earth) may result in not-permissible voltages impairing
or destroying the device or another device connected.

4.  Whenever there may be a risk whatsoever involved in running it, the device has to be switched off immediately and

to be marked accordingly to avoid re-starting. Operator safety may be a risk if:
- there is visible damage to the device or the device is not working as specified
- the device has been stored under unsuitable conditions for a longer time
In case of doubt, please return device to manufacturer for repair or maintenance.

5. 

Warning: 

Do not use these product as safety or emergency stop device, or in any other application where failure of

the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material damage.

1.2  Operating And Maintenance

  If the symbol 

"LO BAT” 

is displayed at the left side of display, the battery is weak, measuring can be continued for

a short  period. If  “bAt” is displayed in the main display the battery is used up and needs to be replaced. Measuring
is no more possible.

  The battery has to be removed, when storing device above 50°C.

Hint: We recommend removing the battery if device is not used for a longer period of time! Risk of Leakage

  Treat device and probes carefully. Use only in accordance with above specification. (do not throw, hit against etc.).

Protect from soiling, especially around the needles.

  After switching on the instrument a segment test (all segments) is displayed, followed by the measuring unit and, if

activated, the auto power off function (please refer to configuration)

1.3 Disposal 

Notice

  Dispense exhausted batteries at destined gathering places.

  Send the device directly to us, if it should be disposed. We will dispose the device appropriate and non-polluting.

1.4 Display 

Elements

1. BAT

Indicates low battery

2. measurement
    display

consider additional arrows and symbols in
display!

3. unit arrows

TDS-measurement in mg/l or conductivity
measurement in µS/cm or mS/cm or Salinity

4. nLF

Non linear temperature compensation is active

5. HLD

Measure value is ‘frozen‘ (Key 3)

1.5 Pushbuttons

key 1:

On/Off key

Press long: off;
Press short: Show temperature

key 2:

mode

  Choose between measuring value / min value / max value
  press 2s: reset min and max value

Key 3:

hold:

 Hold current measuring value ('HLD' in display)

Summary of Contents for GLF-100

Page 1: ...UCTIVITY MEASURING 3 2 1 THE CONDUCTIVITY MEASURING CELL 3 2 2 MEASURING HINTS 3 2 3 TEMPERATURE COMPENSATION 3 2 4 MIN MAX VALUE MEMORY 3 2 5 HOLD FUNCTION 3 3 CONFIGURATION OF THE INSTRUMENT 4 4 ADJUSTING OF THE INSTRUMENT 5 5 SYSTEM MESSAGES 5 6 INSPECTION OF THE ACCURACY ADJUSTMENT SERVICES 5 7 SPECIFICATION 6 precise measurements of conductivity TDS total dissolved solids and Salinity Applica...

Page 2: ...ergency stop device or in any other application where failure of the product could result in personal injury or material damage Failure to comply with these instructions could result in death or serious injury and material damage 1 2 Operating And Maintenance If the symbol LO BAT is displayed at the left side of display the battery is weak measuring can be continued for a short period If bAt is di...

Page 3: ...hem useless very fast Measuring procedure Especially when measuring low conductivity Before immersion to the measuring solution rinse the electrode with deionised water dry with a paper towel and shake remaining water out The measuring is speeded up considerably if the electrode is immersed and pulled out the solution several times Especially when measuring low conductivity the electrode needs suf...

Page 4: ...ng of TDS 0 2000 mg l 50 0 SAL Measuring of Salinity 0 0 50 0 g kg Unit of temperature display C temperature is displayed in C factory setting F temperature is displayed in F temperature compensation only for µS cm and mS cm p r t note below oFF No temperature compensation of conductivity measuring nLF non linear temperature compensation for natural water acc to EN27888 DIN 38404 to measure ground...

Page 5: ...emperature 2 0 2 0 C 3 6 3 6 F Offset setting of temperature measurement see below Scale of temperature 5 00 5 00 Scale adjustment of temperature measurement Display measured value offset 1 scale adjustment 100 Pressing again stores the settings the instruments restarts segment test Please note If there is no key pressed within the menu mode within 60 seconds the adjusting will be cancelled the en...

Page 6: ... 100 C Relative humidity instrument 0 95 r F non condensing Storage temperature 5 50 C Housing Dimension 110 x 67 x 30 mm L x B x D impact resistant ABS membrane keyboard transparent panel Front side IP65 Weight approx 155 g Power Supply 9V Battery type IEC 6F22 included Power Consumption 1 5mA Display ca 11 mm high 4 digit LCD display with additional segments Pushbuttons 3 membrane keys for on of...

Reviews: