 Operating manual G 7500 Series 

page 10 of 34 

 

 

 

 

 

4.2  Status line 

The status line is the first line in the display. 
 

Display 

Meaning 

 

Current time 

If time is flashing, time must be reseted 

Internal memory error. When this error won’t 
disappear after restarting the device, it should 
be returned for repairing. The device will go to a 
safe fallback mode were device settings are 
stored and recalled from the mass storage 
memory if this is available. 

Saving to the internal mass storage memory 
takes longer than intended. When the S will be 
shown permanently in the display, please let 
Windows check the mass storage device for 
errors. When the S still shows up, the device 
needs to be send back for repairing. 

Alarm of a channel is active 

USB 

USB connection has been established. 

LOG 

Logger is active 

BAT 

Rechargeable battery capacity critically low 

Battery indicator 

If the battery capacity flashes, the battery is 
being charged 

4.3  Display elements 

Views can be changed by the right and left keys of the directional pad. With the keys 
up and down the channel can be changed. 

Table view 

Chart view 

Large view 

 

 

 

All channels in one row 

Chart of one channel 

Display of one channel 
with specific parameters 
(here: electrode quality 
and temperature 
compensation)  

Summary of Contents for G 7500 Series

Page 1: ...Operating Manual Water analysis hand held meter G 7500 series Keep for later use www ghm group de Companies Brands of the GHM Englisch...

Page 2: ...UCTIVITY 15 6 3 CONDUCTIVITY MEASUREMENT 15 6 4 ELECTRODES MEASURING CELLS 15 6 5 TEMPERATURE COMPENSATION 16 6 6 DESIGN OF THE SENSOR GWO 5610 17 6 7 COMMISSIONING FILLING OF THE SENSOR GWO 5610 19 6...

Page 3: ...g due to misinterpretation of measured values the operator must have further knowledge in case of doubt the user is liable for any harm damage resulting from misinterpretation due to insufficient know...

Page 4: ...in of this chemical This Symbol indicates irritant substances that can cause inflammation on short term prolonged or repeated contact with the skin or mucous membranes Attention This symbol warns of p...

Page 5: ...electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when u...

Page 6: ...specification sheet of chemicals 3 Consider any statutory provisions guidelines and guidelines of chemicals when disposing This is also for accidentally spilled chemicals dried residues soiled rags o...

Page 7: ...s a deep discharge protection when the batteries reach a limit value the device can no longer be turned on 2 Treat device and electrodes carefully Use only in accordance with above specification do no...

Page 8: ...ang it on a belt Pull the grip labelled open in order to fold it out to a 90 angle Now the product can be positioned on a stable surface Pull the grip labelled open again in order to fold it out to a...

Page 9: ...isplay directly above the soft keys are describing the soft keys function e g back for F1 and change for F2 power button Will turn the device on only when rechargeable batteries are not empty or off o...

Page 10: ...en the S will be shown permanently in the display please let Windows check the mass storage device for errors When the S still shows up the device needs to be send back for repairing A Alarm of a chan...

Page 11: ...alkaline behavior of an aqueous solution pH values below 7 are acidic smaller values indicate higher acidity and values above 7 are alkaline pH 7 neutral The pH measurement is very precise but also s...

Page 12: ...measured can be designed in different ways Clogging or soiling of the diaphragm is a frequent cause of a malfunctioning or sluggish electrode Always handle the glass membrane with extreme care The so...

Page 13: ...used for small sample amounts It comprises a glass electrode with two ceramic diaphragms and liquid electrolyte GE 104 BNC is preferably used for measurements in low ionic media such as rainwater aqua...

Page 14: ...rnail or cloth and is therefore does not constitute a defect or cause for complaint Dirty electrodes must be cleaned The suitable cleaning agents for the pH glass membrane are listed in the table belo...

Page 15: ...500 0 S cm Within the integrated automatic range selection the range with the best resolution is automatically selected 6 4 Electrodes measuring cells 6 4 1 Design and selection Basically there are t...

Page 16: ...ording to EN 27888 is sufficiently accurate Recommended application range of nLF compensation between 60 S cm and 1000 S cm 6 5 2 Linear temperature compensation If the actual function needed for exac...

Page 17: ...ular when it is stored in dry air Therefore it should be checked and maintained regularly and replaced as necessary The electrode contains potassium hydroxide This causes burns All contact with the sk...

Page 18: ...lyte must be filled or added for the initial commissioning of a sensor which is delivered dry when performing maintenance or after use at high temperatures 6 6 3 Service life The sensor signal deterio...

Page 19: ...platinum cup A rapid or excessive formation of lead carbonate after commissioning is an indication of air in the sensor This is usually due to incomplete filling or a leak due to improper fitting of t...

Page 20: ...emove filling screw and top up lost electrolyte e g using disposable syringe 4 Put back and tighten filling screw 5 Top up diaphragm head with electrolyte avoiding air bubbles and place on table cover...

Page 21: ...es the oxygen concentration mg l the oxygen saturation from the electrode signal and the temperature According to DIN38408 C22 all measurements refer to steam saturated air 6 10 Ambient pressure and m...

Page 22: ...ly pressing the Up button and the Down button you can scroll through the parameters Select the parameter you would like to configure 4 When you have selected the desired parameter select them with the...

Page 23: ...cember 2020 at 19 11 The CSV files are then located in the folder DATA 20201231 1911 Single shot logger data is always stored inside the HISTORY directory Calibration data are stored like logger data...

Page 24: ...e logger is running the F1 key will be entitled with stop and can be stopped by pressing the F1 key 7 1 4 3 on keypress With on keypress one measuring will be added to the files inside the HISTORY dir...

Page 25: ...the temperature is measured from the reference channel 7 2 2 reference channel Here the cannel for the ATC can be selected 7 2 2 1 banana sockets Reference temperature measured from RTD on banana soc...

Page 26: ...For calibration the standard buffers for the GMH handheld instruments are used These are pH 4 red pH 7 green und pH 10 blue 7 2 3 2 DIN standard For calibration DIN standard buffers are used pH 1 680...

Page 27: ...k When other electrodes are used the setting has to be changed manually Pt1000 is normally the correct setting for temperature pH as it s standard electrode is the GE 125 When using different electro...

Page 28: ...e are preconfigured with the electrode s cell factor 7 5 2 auto range Automatic range selection of the conductivity measurement is active auto range 0 S cm 500 mS cm 7 5 3 range Manual range selection...

Page 29: ...tarting wait shortly 100000017 data pointer error restart device 100000018 data invalid restart device 100000020 recording stopped logger has been stopped 100000021 recording started logger has been s...

Page 30: ...ers to determine the temperature of the buffer solutions if no GE 125 or similar is used Start the calibration process see above Follow the instructions of the device Clean the electrode and immerse i...

Page 31: ...g a deviation of the measuring device DAkkS calibration certificates are accompanied with a list of individual measurements before and after the adjustment documentation and if applicable graphic repr...

Page 32: ...ected Insert device into silicone cover 12 Reshipment and disposal 12 1 Reshipment All devices returned to the manufacturer have to be free of any residual of measuring media and other hazardous subst...

Page 33: ...5 FS temperature compensation automatic 0 0 50 0 C salinity correction off 0 70 g kg accuracy requirements Ta Tm 20 C 100 sat O2 direct flow 20 cm s conductivity suggested probe LF 425 with Pt 1000 t...

Page 34: ...support and retaining clip dimension 160 x 86 x 37 mm H x B x T incl silicone cover weight approx 300 g incl rechargeable batteries scope of supply device rechargeable batteries micro USB cable and sh...

Reviews: