H87.0.0X.6C-01
Operating manual G 7500 Series
page 29 of 34
8
Error codes (in dataset)
While logging data or querying values via interface, the error codes will not be shown
as human readable text. This is because the relation between error and error
message would get lost between different languages.
Error code
Text
Hint
0
OK
no error
100000000
measuring range overrun
check calibration and sensor
100000001
measuring range underrun
check calibration and sensor
100000010
calculation failed
check settings
100000011
system error
restart device*
100000012
battery empty
recharge device
100000013
no sensor
connect sensor
100000014
recording error
restart device*
100000015
EEPROM checksum wrong
restart device*
100000016
system restarted
device is restarting, wait shortly
100000017
data pointer error
restart device*
100000018
data invalid
restart device*
100000020
recording stopped
logger has been stopped
100000021
recording started
logger has been started
100000022
channel deactivated
channel is deactivated via device
settings
100000023
temp. channel deactivated
check temperature compensation
settings,
activate O2/cond. channel
100000024
no temp. sensor
plug in temperature sensor check
temperature compensation
settings,
activate O2/cond. channel
100000025
no data available
device has not measured anything
-23
sensor module not responding
restart device*
-10
not existing
restart device*
-255
unexpected error
restart device*
-100
calibration error
try to calibrate again
-75
not found
restart device*
-101
not calibrated
perform a calibration
-253
Value not stable
Ensure a stable environment
-251
Not in the temperature range
Check temperature
*Resend device for service, when the error does not disappear