9.3  System messages 

 

°C

%

 

 

BATTERY

• display of the  BAT-arrow left hand down: 

The 

EASY

LOG

 battery is almost spent and needs to be replaced. 

The data recording is still active. 

• permanent display 

BAT

  The battery is spent and the recording has been automatically 

stopped (even if at EASYBus-operation).  

It is still partly possible to read the measured values and to  
read-out the saved logger data. 

 

=> Please return the logger to the manufacturer. 

 

ERROR 

7

The 

EASY

LOG

 has detected a system error. 

• remedy:   Reset the error via GSOFT 40K. 

  If the error message is displayed furthermore please  
  send the logger to the manufacturer for repair.

 

 

 

9.4  Alarm and error messages 

The following messages correspond to a particular measuring channel and are dis-
played accordingly. 

 

ALARM LOW

The measured value is below the selected min-alarm limit. 

This and the current value is displayed alternately.  

 

ALARM HIGH

The measured value is above the selected max-alarm limit.

 

This and the current value is displayed alternately.

 

 

ERROR 

1

The measuring range of this channel is exceeded. 

 

ERROR 

2

The measuring range of this channel is under-run. 

 

ERROR 

9

The sensor for this channel provides invalid values. 

 

ERROR 

11

The variable could not be calculated, because the measuring  
range has been exceeded or a measured value, that is needed for 
the calculation, is not available or faulty. 
 

Note: 

A description of the possible reasons for that error messages is in 
chapter 12 (FAQ).  

 

 

Summary of Contents for 602773

Page 1: ...electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 phone 49 9402 9383 0 fax 49 9402 9383 33 eMail info greisinger de ab f from Version V1 8 Operating Manual EASYLOG 80CL Data logger for humidity...

Page 2: ...logger EASYLOG 80CL is designed for long term monitoring of climate data Not only relative humidity temperature and air pressure can be displayed but also wet bulb tem perature dew point temperature...

Page 3: ...tentials 4 If there is any risk whatsoever involved in running it the device has to be switched off immediately and to be marked accordingly to avoid re starting Operator safety may be at risk if ther...

Page 4: ...play element for max min Logg indicates that logger is recording BAT signals weak battery see chapter 9 AL signals an alarm state at minimum one channel SL signals an active sea level correction for t...

Page 5: ...the display element MAX Note You can immediately switch to the next measuring channel pushing the button again Delete max value Push button for about 5 seconds CLr is displayed and the max values of...

Page 6: ...tart recording Push button for about 5 seconds The main display shows run The secondary display shows no To start recording select YES with the arrow keys button or and verify with Note If no key is p...

Page 7: ...ed if the logger is stopped logger without logger data 7 4 Operating possibilities via software EBxKonfig or EASYBus Configurator The software EBxKonfig V3 9 or EASYBus Configurator V2 0 is needed to...

Page 8: ...ow AL signals an alarm state at minimum one channel The arrow SL signals an active sea level correction 9 2 Logger status messages At stand alone operation no EASYBus connected logger status messages...

Page 9: ...er to the manufacturer for repair 9 4 Alarm and error messages The following messages correspond to a particular measuring channel and are dis played accordingly ALARM LOW The measured value is below...

Page 10: ...at range 10 90 RH Air pressure 1 0 mbar typ 2 5 mbar max Display two 4 digit LCD displays about 7 and 5 8 mm high Interface EASYBus Busload equals to 2 EASYBus elements Battery life time depending on...

Page 11: ...ng transparent screen made of polycarbonate splash proof acc to IP65 excl protection cap Sensor tube about 15 mm made of polyamide Protection cap removable twist plastic head made of polycarbonate IP4...

Page 12: ...oftware delete data Problem A error message is displayed Error message Possible reason procedure remedy measuring range exceeded Measuring range of the chan nel has been exceeded Operate logger within...

Page 13: ...tion diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be connected to terminal 11 of the GNR NPN to be connected to terminal 10...

Page 14: ...E40 0 01 6C 04 Operating Manual EASYLOG 80CL Page 14 of 14...

Reviews: