GREISINGER electronic T-Logg 160 Operating Manual Download Page 4

E39.0.31.6C-03 

Operating instruction T-Logg 160 

page 4 of 8

 

 

 

BATTERY

 

Changing display

 

BAT

 / Value: 

  The battery of the 

T-Logg

 is getting discharged soon. 

  Replace the battery briefly. Even so data logging is still active. 
 
• Constant display 

BAT

  The Logger battery is empty now and data logging has been stopped. 
  Replace the battery immediately. 
 

 

ALARM LOW

The measured value (unit arrow points to the concerned measuring) is below the  
min. alarm limit. 
 

 

ALARM HIGH

The measured value (unit arrow points to the concerned measuring) has exceeded the  
max. alarm limit. 
 

 

ERROR 1

The measured value has exceeded the measuring range of the logger. 
 
 

 

ERROR 2

The measured value has been fallen below the measuring range of the logger. 
 
 

 

ERROR 7

The 

T-Logg

 has detected a system fault. 

 Remedy: 

  Remove battery and wait about 30 minutes. Then insert the battery again. 

 If the error message displayed furthermore, please send the logger to the manufacturer 

  to repair. 

 

7  Battery service life and recording time 

 
 

Measuring cycle:

 

4 sec. 

10 sec. 

15 min. 

5 hours 

 

Recording time:

 

17.8 hours 

44 hours 

166 days 

Approx. 9 years 

 

Battery service life:

 

   

—    

Approx. 8 month  Approx. 3 years 

   

—    

 

 

Please note!

 

Short measuring cycles as well as frequently measuring data transfer result in a reduction of 
the battery service life! 
 
Even if the T-

Logg is connected, power consumption is increased. So it’s important to 

connect the logger with the USB interface of the PC only as long as necessary! 
 

 

Summary of Contents for T-Logg 160

Page 1: ...E39 0 31 6C 03 GREISINGER electronic GmbH Data logger for humidity temperature as of version V1 0 Operating Manual T Logg 160...

Page 2: ...Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification To protect the battery...

Page 3: ...layed as well STOP The T Logg is stopped No data are recorded The logger memory is empty The logger is reset and can be restarted HALT The T Logg has been halted The stored data can be read The logger...

Page 4: ...the logger ERROR 2 The measured value has been fallen below the measuring range of the logger ERROR 7 The T Logg has detected a system fault Remedy Remove battery and wait about 30 minutes Then inser...

Page 5: ...nt to replace the battery on yourself it is also possible to send the device to us We will replace the battery cost effectively and appropriately Replacement 1 Remove the 4 screws at the front of the...

Page 6: ...ector again and move it back into the probe tube 10 Plug the sensor element to the connector mind right mounting position Note The connection pins can break if they are inappropriately handled 11 Scre...

Page 7: ...changeable Battery service life approx 3 years if interval is 15 min depending on measuring cycle and operating temperature Interface Serial interface 3 pin miniature plug Data communication via inter...

Page 8: ...E39 0 31 6C 03 Operating instruction T Logg 160 page 8 of 8...

Reviews: