1  General 

The logger 

T-Logg 160 

is designed as a cost-efficient solution for monitoring humidity and temperatures. It enables 

an individual programming of the recording time. The last 16,000 measuring values can be stored in the memory.  
In addition, the LCD-display indicates both the temperature measured at the moment and the operating status of the 
logger. 

2  Required accessory 

The USB interface of your PC is used to program, start and read out the 

T-Logg

For this following accessory is required: 

  Interface converter 

USB 100 

or

 USB 100 SL

 for direct connection to the USB port of the PC, 

cable length approx. 1m. 

 

MINISOFT

 software (version 7.18 or later / free of charge) to start the logger and read out the logger data. 

Notice: 

It’s also possible to use the T-Logg with the comprehensive software 

GSOFT 40K 

  

(version 7.18 or later).

 

3  Safety instructions 

This device has been designed and tested in accordance with the safety regulations for electronic devices. 
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and 
special safety advises given in this manual will be adhered to when using the device. 

1.  Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any 

other climatic conditions than those stated under ‘Specification’ 

To protect the battery the max. permissible storage and transport temperature of the device is +85°C.

 

2.  General instructions and safety regulations for electric, light and heavy current plants, including domestic safety 

regulations (e.g. VDE), have to be observed. 

3.  If device is to be connected to other devices (e.g. via PC) the circuitry has to be designed most carefully. Internal 

connection in third party devices (e.g. connection GND and earth) may result in not-permissible voltages 
impairing or destroying the device or another device connected. 

4.  If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be 

marked accordingly to avoid re-starting. 
Operator safety may be a risk if: 
-  there is visible damage to the device. 
-  the device is not working as specified. 
-  the device has been stored under unsuitable conditions for a longer time. 
In case of doubt, please return device to manufacturer for repair or maintenance. 

5.  Warning:

 

Do not use this product as safety or emergency stop devices, or in any other application where failure of the 
product could result in personal injury or material damage. 

 

Failure to comply with these instructions could result in death or serious injury and material damage! 

 
 

Summary of Contents for T-Logg 160

Page 1: ...E39 0 31 6C 03 GREISINGER electronic GmbH Data logger for humidity temperature as of version V1 0 Operating Manual T Logg 160...

Page 2: ...Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification To protect the battery...

Page 3: ...layed as well STOP The T Logg is stopped No data are recorded The logger memory is empty The logger is reset and can be restarted HALT The T Logg has been halted The stored data can be read The logger...

Page 4: ...the logger ERROR 2 The measured value has been fallen below the measuring range of the logger ERROR 7 The T Logg has detected a system fault Remedy Remove battery and wait about 30 minutes Then inser...

Page 5: ...nt to replace the battery on yourself it is also possible to send the device to us We will replace the battery cost effectively and appropriately Replacement 1 Remove the 4 screws at the front of the...

Page 6: ...ector again and move it back into the probe tube 10 Plug the sensor element to the connector mind right mounting position Note The connection pins can break if they are inappropriately handled 11 Scre...

Page 7: ...changeable Battery service life approx 3 years if interval is 15 min depending on measuring cycle and operating temperature Interface Serial interface 3 pin miniature plug Data communication via inter...

Page 8: ...E39 0 31 6C 03 Operating instruction T Logg 160 page 8 of 8...