E39.0.31.6C-03
Operating instruction T-Logg 160
page 3 of 8
4 Connection
■
Assignment of the input jack
(front view to the pins):
Interface [
–]
Not connected
Interface [+]
5 Advice regarding state of logger upon delivery:
Upon its delivery the logger is in a kind of 'sleeping state': The display does not show anything, the power
consumption is at its minimum.
The
T-Logg
‘wakes up’ as a communication link with a software has been established. After that the message
‘Stop‘ appears at display. The logger is ready for operation now.
6 Operating mode display:
The
T-Logg 100
is equipped with a 10 mm LCD display.
The main purpose of the LCD display is to indicate the temperature and humidity.
Depending on the operating mode of the logger, other messages will be displayed as well.
STOP
:
The
T-Logg
is ‘stopped’. No data are recorded. The logger memory is empty.
The logger is reset and can be restarted.
HALT
:
The
T-Logg
has been ‘halted’. The stored data can be read.
The logger memory is not empty.
DISPLAY OF TEMPERATURE
:
The small unit arrow points to °C. The logger is active. Measurements are carried out at certain
intervals and will be stored. The temperature value will be displayed.
DISPLAY OF HUMIDITY
:
The small unit arrow points to %. The logger is active. Measurements are carried out at certain
intervals and will be stored. The humidity value will be displayed.
START DELAY
:
The logger is active, but no data are recorded. As soon as the start delay time has expired, the
logger will start recording in accordance with the starting conditions programmed before.
START ALARM
:
The logger is active, but no data are recorded. Recording will start as soon as the temperature and
humidity are within the min. and max. alarm limits.