5 Configuration

In the configuration the devices parameters can be changed.
The jumper has to be set, p.r.t. figure rigthand side. To set or
remove jumper, the housing cover has to be removed. Ex works
the jumper is set.
To change parameters press “SET” (key 1) two seconds, then the
parameter selection is started with the first parameter (display
shows “CELL”).
By pressing “SET” the desired parameter is selected, the editing of
the parameter values happens via keys 

5

 (key 3) or 

6

 (key 2).

Pressing “SET” again (after editing the parameter values) returns
to the parameter selection
Pressing “SET” again after the last parameter finishes the
configuration, stores the changes and restarts the device.

5.1  'CELL': Setting of the cell-constant of the connected measuring cell

Possible inputs: 0.300...1.200. Unit is 1/cm.
The value was evaluated and stored ex works for the connected electrode and is in between 0.45...0.65.

5.2  'Func': Setting of the measuring function

Possible inputs:

cond  conductivity
reSi:

resistivity

tdS:

TDS (total dissolved solids

SAL: Salinity = salt content of sea water

5.3  'rAnG': Setting of the conductivity or resistivity range

Choice the desired measuring range (range is depending on the function set above).
Possible inputs:

Range (rAnG)=0

Range (rAnG)=1

Range (rAnG)=2

Range (rAnG)=3

Conductivity (cond) 0.0 ... 200.0 mS/cm

0.00 ... 20.00 mS/cm 0 ... 2000 µS/cm

0.0 ... 200.0 µS/cm

Resistivity (rESi)

5.0 ... 100.0 Ohm·cm 50 ... 1000 Ohm·cm

0.50 ... 10.00 kOhm·cm 5.0 ... 100.0 kOhm·cm

TDS (tdS)

-

-

0 ... 2000 mg/l

0.0 ... 200.0 mg/l

Salinity (SAL)

0.0 ... 70.0

-

-

-

5.4  't.Cor': Setting of temperature compensation method

Possible inputs: off:  no temperature compensation

nLF:non linear compensation for natural water according to EN27888 (DIN 38404).

 

For measurings of ground and surface water, drinking water or pure water.

 

Attention: restricted temperature range: -5..105°C

 

Lin: linear temp. compensation (for other aqueous solutions, correct factor has to be

 entered)

5.5  't.Lin': Setting of temperature coefficient (only for t.Cor = Lin)

Possible input: 0.300 .. 3.000 [%]

5.6  't.rEF': Setting of reference temperature (only for t.Cor = nLF or Lin)

Possible input: 20°C or 25°C

5.7  'C.tdS': Setting of TDS-factor

Calculation factor for TDS (Total Dissolved Solids)-measurements, possible input: 0.40 .. 1.00.
The calculation factor depends on the composition of the medium and has to be determined
for each type of water.

If the jumper is removed from
the shown contacts, the
configuration is inaccessible,
values are protected.

Summary of Contents for GLMU 200 MP

Page 1: ...3 wire output signal electrically isolated Auxiliary energy supply voltage Uv 12 30 V DC 4 20mA Uv 12 30 V DC max 10mA 0 1V Uv 18 30 V DC max 10mA 0 10V or refer to type plate Reverse voltage protect...

Page 2: ...cluding domestic safety regulations e g VDE have to be observed 3 If device is to be connected to other devices e g via PC the circuitry has to be designed most carefully Internal connection in third...

Page 3: ...t or voltage output is set by works and cannot be changed 3 4 Terminal assignment of meas cell socket contact no Standard Cell K 0 55 1 Temperature 2 Temperature GND 3 Supply 1 4 Signal 1 5 Signal 2 6...

Page 4: ...linity salt content of seawater can be determined in the measuring mode SAL basis International Oceanographic Tables IOT The salinity of standard seawater is 35 35g salt per 1kg seawater The values ar...

Page 5: ...al Temperature below 5 C not allowed Choose right measuring range for conductivity Err 7 System fault Error in device Disconnect from supply and reconnect If error remains return to manufacturer Err 9...

Page 6: ...the conductivity or resistivity range Choice the desired measuring range range is depending on the function set above Possible inputs Range rAnG 0 Range rAnG 1 Range rAnG 2 Range rAnG 3 Conductivity...

Page 7: ...5 0 C or oFF offset is deactivated 0 0 ex works 5 12 SCAL with Temp arrow Scale correction of temperature measuring The scale of the measuring is changed by this value Calculation see below Max input...

Reviews: