GREISINGER electronic GLMU 200 MP Operating Manual Download Page 5

T14.0.0X.6C-02

User’s Manual GLMU 200 MP

page 5 of 8

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

4.3  Error and system messages

Display Description

Possible fault cause

Remedy

Err.1

measuring range
exceeded

Wrong signal

Temperature above 140°C not allowed.
Choose right measuring range for
conductivity

Err.2

Measuring value below
measuring range

Wrong signal

Temperature below -5°C not allowed.
Choose right measuring range for
conductivity

Err.7

System fault

Error in device

Disconnect from supply and reconnect.
If error remains: return to manufacturer

Err.9

Sensor error

Sensor or cable defective Check sensor, cable and connections

Er.11

Calculation not possible

Calculation variable
missing or invalid

Check temperature

8.8.8.8

Segment test

The transducer performs a display test for 2 seconds after power up.
After that it will change to the display of the measuring.

Summary of Contents for GLMU 200 MP

Page 1: ...3 wire output signal electrically isolated Auxiliary energy supply voltage Uv 12 30 V DC 4 20mA Uv 12 30 V DC max 10mA 0 1V Uv 18 30 V DC max 10mA 0 10V or refer to type plate Reverse voltage protect...

Page 2: ...cluding domestic safety regulations e g VDE have to be observed 3 If device is to be connected to other devices e g via PC the circuitry has to be designed most carefully Internal connection in third...

Page 3: ...t or voltage output is set by works and cannot be changed 3 4 Terminal assignment of meas cell socket contact no Standard Cell K 0 55 1 Temperature 2 Temperature GND 3 Supply 1 4 Signal 1 5 Signal 2 6...

Page 4: ...linity salt content of seawater can be determined in the measuring mode SAL basis International Oceanographic Tables IOT The salinity of standard seawater is 35 35g salt per 1kg seawater The values ar...

Page 5: ...al Temperature below 5 C not allowed Choose right measuring range for conductivity Err 7 System fault Error in device Disconnect from supply and reconnect If error remains return to manufacturer Err 9...

Page 6: ...the conductivity or resistivity range Choice the desired measuring range range is depending on the function set above Possible inputs Range rAnG 0 Range rAnG 1 Range rAnG 2 Range rAnG 3 Conductivity...

Page 7: ...5 0 C or oFF offset is deactivated 0 0 ex works 5 12 SCAL with Temp arrow Scale correction of temperature measuring The scale of the measuring is changed by this value Calculation see below Max input...

Reviews: