4.3 Error and system messages
Display Description
Possible fault cause
Remedy
Err.1
measuring range
exceeded
Wrong signal
Temperature above 140°C not allowed.
Choose right measuring range for
conductivity
Err.2
Measuring value below
measuring range
Wrong signal
Temperature below -5°C not allowed.
Choose right measuring range for
conductivity
Err.7
System fault
Error in device
Disconnect from supply and reconnect.
If error remains: return to manufacturer
Err.9
Sensor error
Sensor or cable defective Check sensor, cable and connections
Er.11
Calculation not possible
Calculation variable
missing or invalid
Check temperature
8.8.8.8
Segment test
The transducer performs a display test for 2 seconds after power up.
After that it will change to the display of the measuring.