Arrow to Temp indicates
temperature display

4 Operation

4.1  General informations about the conductivity measuring

4.1.1  The standard measurung cell

During the measurement, the conductivity probe must be be dipped at least 30mm into the medium.
At the one hand to allow a reproducable electrical field around the measuring cell, at the other hand to allow
the temperature sensor (located in metall cover within the measuring cell) to adjust to the medium
temperature.
When changing between different solution the probe should be rinsed thoroughly.

Measuring cell must never come into contact with water-repellent materials such as oil or silicone!

If conductivity measured is much higher or lower than expected this may be due to the electrode being soiled
with nonconducting or conducting residues. In this case the measuring cell has to be cleaned with a watery
soap solution.
Especially when measuring media with low conductivities the electrode the media should flow at sufficient
speed though the cell.

4.1.2 Salinity 

measuring

The salinity (salt content) of seawater can be determined in the measuring mode „SAL“ (basis: International
Oceanographic Tables; IOT)
The salinity of standard-seawater is 35 ‰ ( 35g salt per 1kg seawater).
The values are displayed in ‰ (g/kg).

4.1.3 TDS 

measuring

At the TDS measuring the dry residues or 

t

otal 

d

issolved 

s

olids value is caculated by means of the

conductivity and a calculation factor (C.tdS). This factor has to be evaluated for the refering type of solution.
The values are displayed in mg/l.

4.2 Display functions

4.2.1  Currently measured values

During normal operation 

the conductivity display value 

is displayed.

By short press of key ‘SET’ (1) the 

temperature

 in 

[°C]

 will displayed vor 10 seconds.

display of conductivity value

display temperature

4.2.2  Min/Max Value Memory

watch Min values (Lo):

press ‘down‘(2) shortly once

display changes between ‘Lo‘ and Min values

watch Max values (Hi):

press ‘up‘(3) shortly once 

display changes between ‘Hi‘ and Max values

restore current values:

press ‘down‘(2) or ’up’(3) once again

current values are displayed

clear Min-values: 

press ‘down’(2) for 2 seconds

Min values are cleared. The display shows shortly ‘CLr‘.

clear Max-values: 

press ‘up’(3) for 2 seconds

Max values are cleared. The display shows shortly ‘CLr‘.

After 10 seconds the currently measured values will be displayed again.

Summary of Contents for GLMU 200 MP

Page 1: ...3 wire output signal electrically isolated Auxiliary energy supply voltage Uv 12 30 V DC 4 20mA Uv 12 30 V DC max 10mA 0 1V Uv 18 30 V DC max 10mA 0 10V or refer to type plate Reverse voltage protect...

Page 2: ...cluding domestic safety regulations e g VDE have to be observed 3 If device is to be connected to other devices e g via PC the circuitry has to be designed most carefully Internal connection in third...

Page 3: ...t or voltage output is set by works and cannot be changed 3 4 Terminal assignment of meas cell socket contact no Standard Cell K 0 55 1 Temperature 2 Temperature GND 3 Supply 1 4 Signal 1 5 Signal 2 6...

Page 4: ...linity salt content of seawater can be determined in the measuring mode SAL basis International Oceanographic Tables IOT The salinity of standard seawater is 35 35g salt per 1kg seawater The values ar...

Page 5: ...al Temperature below 5 C not allowed Choose right measuring range for conductivity Err 7 System fault Error in device Disconnect from supply and reconnect If error remains return to manufacturer Err 9...

Page 6: ...the conductivity or resistivity range Choice the desired measuring range range is depending on the function set above Possible inputs Range rAnG 0 Range rAnG 1 Range rAnG 2 Range rAnG 3 Conductivity...

Page 7: ...5 0 C or oFF offset is deactivated 0 0 ex works 5 12 SCAL with Temp arrow Scale correction of temperature measuring The scale of the measuring is changed by this value Calculation see below Max input...

Reviews: