12
Mini-Convectron
®
Module with DeviceNet
®
Digital Interface
Chapter 3
3.8
Mounting Flange
The
275
Mini-Convectron
Vacuum Gauge Module can be equipped with a
metal seal, O-ring, or VCR type mounting flange as listed in Table 3-1. The
type of flange must be specified when ordering the
275
Mini-Convectron
Vacuum Gauge Module.
Table 3-7
DeviceNet Network (NET) LED States
Network State
LED State
Description
Not Powered, Not On-
Line
OFF
Module is not on-line
• The module has not completed the DUP_MAC_ID test yet.
• The module may not be powered, look at module status LED.
Self Test
Flashing
Red/Green
Module is in self-test.
On-line, Not Connected
Flashing Green
Module has passed the Dup_MAC_ID test, is on-line, but has no
established connections to other nodes. It means that the module has
no established connections.
On-line, Connected
Green
The module is allocated to a Master. Device is operating normally.
Critical Link Failure
Red
Failed communication device. The module has detected an error that
has rendered it incapable of communicating on the network
(Duplicate MAC ID or Bus-off).
Connection Time Out
Flashing Red
The I/O connection is in the timed-out state.