Granville-Phillips Convectron 275 series Instruction Manual Download Page 36

5   Service and Maintenance

36

Series 275 Convectron Gauge Bridge Amp PC Board

Instruction Manual - 275331

5.3.2  Convectron Gauge Calibration

An atmospheric pressure calibration is performed on the Convectron Gauge using N

2

 at the factory 

before the gauge is shipped. The factory calibration sets the atmospheric calibration point for N

2

 to 

760 Torr.

Periodic resets of the vacuum chamber calibration point improve the accuracy and repeatability of 
the Convectron Gauge. Regardless of the process gas being used, always use N

2

 or air to calibrate 

the Convectron Gauge at vacuum chamber pressure.

1.

Using N

2

 or air, pump down the vacuum chamber pressure to a pressure that is lower than 

10

–4

 Torr. When the pressure indication in the 1 x 10

–4

 Torr range has stabilized, a Convectron 

Gauge calibration at vacuum chamber pressure may be performed.

2.

Use a high-precision, high-input impedance (Zin > 1 MW) DVM to measure the voltage across 
the analog output connections (pins #11 & 12 of the interconnect cable - see Figure 3-4).

3.

Use a flat-head instrument screwdriver to adjust the ADJUST VACUUM potentiometer to a 
reading of 0.0 Vdc. The vacuum adjust potentiometer is # R201, shown on Figure 2-2 and 
Figure 3-5.

5.3.3  Convectron Gauge Test Procedure

The small diameter sensor wire inside the gauge tube can be damaged by even small voltages. Do 
not perform electrical continuity tests with instruments applying in excess of 0.1 volt when the 
gauge is at vacuum, or 2 volts when at atmospheric pressure.

1.

Vent the vacuum chamber to atmosphere and allow the gauge to cool to ambient temerature.

2.

Disconnect the cable fromthe gauge.

3.

Use a 

low-voltage (maximum 0.1 V) ohmmeter

 to check resistance values across the pins on 

the gauge. See Figure 5-1.

The 

g

auge should show the resistances values listed in Figure 5-1. Pin numbers are embossed on the 

blue plastic pin feed-through of the gauge. Measure the resistance across the pins with the gauge at 
atmospheric pressure, using an ohmmeter that will not apply more than 10 mA.

4.

Replace the gauge tube if required.

5.

Connect the cable and check the gauge for proper operation.

Figure 5-1

Convectron Gauge Connector and Resistance in Ohms

• Pins 1 to 2: 19 to 22 ohms
• Pins 2 to 3: 50 to 60 ohms
• Pins 1 to 5: 180 to 185 ohms
If the resistance from pins 1 to 2 reads about 800 ohms, 
the sensor wire in the gauge is broken. Replace the 
gauge tube. 

If the resistance values shown here are correct, 
but you still think the gauge is not reading 
correctly, the gold plating on the sensor wire 
may be eroded and the gauge will have to be 
replaced.

Summary of Contents for Convectron 275 series

Page 1: ...Series 275 Instruction Manual Granville Phillips Convectron Vacuum Gauge Bridge Amp PC Board with Analog Output Signal Instruction manual part number 275331 Revision D November 2016...

Page 2: ...All other trademarks and registered trademarks are the properties of their respective owners Customer Service Technical Support MKS Pressure and Vacuum Measurement Solutions MKS Instruments Inc Granvi...

Page 3: ...ier PC Board 17 3 2 1 Install the Interconnect Cable 18 Chapter 4 Operation 21 4 1 Theory of Operation 21 4 1 1 General Precautions 22 4 2 Preparing for Pressure Measurement 22 4 3 Gas Type 22 4 4 Pre...

Page 4: ...4 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 Notes...

Page 5: ...ure to comply with these requirements Read Instructions Read all safety and operating instructions before operating the product Retain Instructions Retain the Safety and Operating Instructions for fut...

Page 6: ...fety features are maintained Do not use this product if it has unauthorized modifications Replacement Parts When replacement parts are required be certain to use the replacement parts that are specifi...

Page 7: ...ill operation where a malfunction might cause high internal pressures provide an appropriate pressure relief device Series 275 Gauges should not be used above 1000 Torr 1333 mbar 133 kPa 19 psi true p...

Page 8: ...Shipping damage on returned products as a result of inadequate packaging is the Buyer s responsibility For Customer Service Technical Support MKS Pressure and Vacuum Measurement Solutions MKS Instrum...

Page 9: ...ty from the date of shipment for new Granville Phillips Products The MKS Instruments Inc General Terms and Conditions of Sale provides the complete and exclusive warranty for MKS Granville Phillips Di...

Page 10: ...1 Safety Instructions 10 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 Notes...

Page 11: ...2 Kovar Kapton gold plated Tungsten and borosilicate glass Table 2 1 Specifications Function Description System Measuring range for Air or N2 See the Operation Chapter of this instruction manual for c...

Page 12: ...NPT pipe thread 1 2 inch compression fitting 1 8 NPT pipe thread 1 2 inch tubulation 6 4 cm 2 52 in VCR type female fitting 1 4 inch VCR type female fitting standard 1 2 inch VCR type female fitting...

Page 13: ...2 Specifications 13 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 2 4 Bridge Amplifier PC Board Figure 2 2 Bridge Amplifier PC Board...

Page 14: ...2 Specifications 14 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 Notes...

Page 15: ...ipe wrench to further tighten the press ring You may point the electrical pins of the gauge tube anywhere you wish in a 360 horizontal circle for optimum routing of the gauge tube cable 1 8 NPT Mount...

Page 16: ...module to the vacuum chamber 1 Attach a metal hose clamp or other metal clamp to the gauge stem 2 Install a 3 31 mm2 12 AWG or larger copper wire between the clamp and a metal ground lug bolt or stud...

Page 17: ...ction Manual 275331 3 2 Install the Bridge Amplifier PC Board Install the bridge amplifier PC board in your existing enclosure Dimensions are given in Figure 3 3 Figure 3 2 Ground Wire Attached to the...

Page 18: ...nect Cable 1 Connect the cable between the bridge amplifier PC board and the Convectron Gauge See Figure 3 4 2 Connect 12 Vdc power to the PCB pins 13 14 and 15 See Figure 3 4 3 Connect the analog out...

Page 19: ...3 Installation 19 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 Figure 3 5 Convectron Gauge Bridge Amplifier Schematic...

Page 20: ...3 Installation 20 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 Notes...

Page 21: ...ungsten The heated sensing wire loses more heat as the ambient gas pressure increases The more molecules contact the sensing wire the more power is required to keep the sensing wire at a constant temp...

Page 22: ...your vacuum system is air or N2 If you are using other gases you must follow the instructions in Section 4 5 You are reasonably familiar with the general theory of operation of thermal conductivity ga...

Page 23: ...r N2 air has approximately the same calibration With proper precautions the Convectron Gauge may be used for pressure measurement of certain other gases NOTE The information in this section applies on...

Page 24: ...true pressure of 2 x 10 2 Torr for CH4 the heat loss from the sensor is the same as at a true pressure of 3 x 10 2 for N2 see Figure 4 2 The curves at higher pressure vary widely from gas to gas becau...

Page 25: ...rom the 10 Torr N2 equivalent indicated pressure to the Helium curve and then move horizontally to the left to reveal a true pressure of 4 5 Torr Thus 4 5 Torr Helium pressure produces an indication o...

Page 26: ...4 Operation 26 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 Figure 4 2 Convectron Gauge Indicated vs True Pressure Curve 10 4 to 10 1 Torr...

Page 27: ...4 Operation 27 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 Figure 4 3 Convectron Gauge Indicated vs True Pressure Curve 10 1 to 1000 Torr...

Page 28: ...4 Operation 28 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 Figure 4 4 Convectron Gauge Indicated vs True Pressure Curve 10 1 to 1000 Torr...

Page 29: ...4 Operation 29 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 Figure 4 5 Convectron Gauge Indicated vs True Pressure Curve 10 4 to 10 1 mbar...

Page 30: ...4 Operation 30 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 Figure 4 6 Convectron Gauge Indicated vs True Pressure Curve 10 1 to 1000 mbar...

Page 31: ...gnal The analog output voltage signal provides a specific voltage at specific pressures as illustrated in Figure 4 8 Use the pin numbers 11 and 12 of the connector on the bridge amplifier PC board see...

Page 32: ...4 Operation 32 Series 275 Convectron Gauge Bridge Amp PC Board Instruction Manual 275331 Figure 4 8 Analog Output Voltage Signal...

Page 33: ...ility For Customer Service Technical Support MKS Pressure and Vacuum Measurement Solutions MKS Instruments Inc Granville Phillips Division 6450 Dry Creek Parkway Longmont Colorado 80503 USA Tel 303 65...

Page 34: ...roduct has been dropped or damaged Replacement Parts When replacement parts are required be certain to use the replacement parts that are specified by Granville Phillips or that have the same characte...

Page 35: ...repair or replace the plumbing 3 Repair the leak or replace the vacum pump 4 Repair or replace the cable Pressure reading is inaccurate 1 The bridge amp PC board is not calibrated for the process gas...

Page 36: ...and Figure 3 5 5 3 3 Convectron Gauge Test Procedure The small diameter sensor wire inside the gauge tube can be damaged by even small voltages Do not perform electrical continuity tests with instrum...

Page 37: ...the sensor 1 Hold the gauge with the main body horizontal and the port projecting upward at an angle of 45 degrees Slowly fill it with solvent using a standard wash bottle with the spout inserted in t...

Page 38: ...r with the Convectron Gauge Attached 1 Bake the process chamber at the desired temperature for the desired time The Convectron gauge tube must not exceed 150 oC 2 Allow the system to cool to ambient t...

Page 39: ...ction manual part number 275173 Revision A August 2012...

Page 40: ...Customer Service Technical Support MKS Pressure and Vacuum Measurement Solutions MKS Instruments Inc Granville Phillips Division 6450 Dry Creek Parkway Longmont Colorado 80503 USA Tel 303 652 4400 Fax...

Reviews: