4 Operation
31
Series 275 Convectron Gauge Bridge Amp PC Board
Instruction Manual - 275331
4.6 Analog Output Voltage Signal
The analog output voltage signal provides a specific voltage at specific pressures as illustrated in
Figure 4-8. Use the pin numbers 11 and 12 of the connector on the bridge amplifier PC board (see
Figure 3-4) to use the voltage signal.
Figure 4-7
Convectron Gauge Indicated vs. True Pressure Curve;
10
-1
to 1000 mbar.