Thermo-shaker PCMT
Version 1.01 - January 2013
Page 5
Operating Instructions
3. Getting started
3.1
Unpacking
Remove packing materials carefully, and retain for future shipment or storage of the unit.
Examine the unit carefully for any damage incurred during transit. The warranty does not
cover in-transit damage.
Attention!
Automatic balancing system in this product produces light metal-like noise when moving
the unit which is likely to be heard during unpacking. It is normal occurrence and does not
indicate a fault.
3.2
The Thermo-shaker PCMT set includes:
• Thermo-shaker PCMT ....................................................................................1 piece
• Microtube block ...............................................................................................1 piece
• Spare rubber belt ..........................................................................................2 pieces
• External Power Supply.....................................................................................1 piece
• Operating instructions; CE Certificate...............................................................1 copy
3.3
Optional blocks
• HC32 block ...................................for 20x0.2 ml micr 12x1.5 ml microtubes
• HC24N block .......................................................................for 24x1.5 ml microtubes
• HC18 block .....................................................for 20x0.5 ml + 12x1.5 ml microtubes
• HC24 Block.........................................................................for 24x2.0 ml microtubes
• HC96 Block ..............................for 96-well PCR microplate (semi-skirted, unskirted)
3.4
Set up:
• Pl a c e the unit upon even horizontal non-flammable surface away from any flammable
materials, remove protective film display.
• To provide optimum ventilation ensure clearance around the device - 20 cm on side
faces.
• Pl u g the external power supply into the socket at the rear side of the unit and position
the unit so that there is easy access to the power switch and external power supply.
3.5
Installing blocks
(if block is not installed at the moment of delivery)! Disconnect from the
mains!).
Connect the plug
to the contact terminal on the underside of the block. Align the block
so that the warning label is facing the front of the unit and secure with the four knurled
screws
.
3.6
Changing blocks
Disconnect the external power supply from the
mains. Remove the four knurled screws
,
disconnect the plug
. Select the new block.